Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 34 of 60
6.5.
Reset (Default Setting)
You have to calibrate the instrument after executing this function.
The device can be restored to the original factory defaults.
Alarm levels
ALARM LO
20 % of measurement range
ALARM HI
50 % of measurement range
Calibration
Standard Calibration
Current Loop
Output
4 ... 20mA
Signal
0
…
20 ppm (Low range)
0
…
100 ppm (Standard range)
0
…
2 000 ppm (High range)
Modbus
Slave ID
10
Mode
RTU
Baud rate
115200
Parity
even
Termination
Enable
AutoPurge
Meas Time
1 min
Purge Time
1 min
Password
Set to default
0000
Response Factor
ISOBUTENE
x 1.00
How To
–
Reset to factory defaults
1.
Go to the Instrument Menu and select RESET.
2.
Select “run” resets the configuration to factory defaults.
Customized response factors that are generated via PC software SPID2 Control Center
will be deleted after reset.