Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 23 of 60
6.2.1.
Calibration Data
The device supports two different methods for calibration. (
Standard Calibration
with SPAN Gas
[Isobutene in air] and
Extended Calibration
with SPAN Gas [User gas in air] with known concentration)
This menu setting allows you to select which type of calibration data are used for measurement.
Both calibration data for
Extended Calibration
and
Standard Calibration
are stored
independently of one another.
How To
–
Select Calibration Data
1.
Go to the Instrument Menu and select MEAS CONFIG
2.
Select “run” to start measurement configuration
3.
Select Calibration, “
change
”
calibration data setting
and “accept”