CLINICIAN
51
Pressure Settings
1. From the Clinician Mode
Home
screen, press
Therapy
,
then
Next
, then
Settings
, then
Pressure
to access this
screen.
2. Use the
+
and
-
buttons to change the
desired pressure. Pressure can be set
from 25 to 200 mmHg in increments of
25 mmHg.
3. Press
Exit
to return to the
Settings
screen.
Default setting is 125 mmHg (ac
/ - 10 mmHg).
Intensity Control
1. From the Clinician Mode
Home
screen, press
Therapy
, then
Next
, then
Settings,
then
Intensity
to change intensity level.
•
Intensity is related to the time it takes to reach the target therapy level
after the initiation of therapy.
•
The lower the intensity setting, the slower the target therapy level will
be reached.
•
It is recommended that new patients begin therapy at the lowest
intensity setting as this allows for slower increase of negative pressure
once the foam is compressed in the wound.
•
The intensity can remain at the minimum setting throughout the entire
length of treatment, if desired.
2. Press to change levels. Green crescent changes with each setting.
Low
Medium
High
Default setting is Low.
Pressure
+
_
125 mmHg
Exit
+
_
Exit
Intensity
Intensity
Intensity
Содержание ACTIV.A.C.
Страница 1: ...User Manual For Patients and Clinicians ACTIV A C THERAPY SYSTEM Rx Only...
Страница 2: ......
Страница 4: ......
Страница 5: ...For Patient Use ACTIV A C THERAPY SYSTEM...
Страница 6: ......
Страница 44: ......
Страница 45: ...For Clinician Use Only Patients Refer to Previous Section of This Manual ACTIV A C THERAPY SYSTEM...
Страница 46: ......
Страница 82: ...CLINICIAN 76...
Страница 83: ...CLINICIAN 77...