Blockage Alert
Low Priority Alert - This alert screen appears when the tubing may be blocked. It is indicated by a single
audible tone.
Therapy will continue during this alert.
This alert is fixed by checking the tubing.
1. Make sure both clamps on the tubing are open.
2. Make sure that the tubing is not kinked, crimped or
blocked in any way.
3. If the Blockage Alert remains after completing steps 1
and 2, lower the therapy unit and tubing to level with
or below the wound site.
4. Press
Exit
to return to the Patient Mode
Home
screen.
Potential Blockage Detected
Blockage
Alert
Blockage
Continuous
0mmHg
Exit
Press ‘?’ for more information
Exit