Battery Critical Alarm
Medium Priority Alarm - This alarm screen appears about 30 minutes before the battery power runs out.
It is indicated by a repeating audible tone.
Press
Audio Pause
to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will continue during this alarm.
However, if the unit is not plugged in
within 30 minutes, therapy will turn off.
This alarm is fixed by recharging the battery.
1. Connect the therapy unit to a wall outlet using the
power supply. Refer to the
Battery Charging
Instructions
section for more information.
2. Press
Reset
to return to the Patient Mode
Home
screen.
3. Make sure therapy is on. The
On / Off
button will be green.
4. Press
On / Off
to restart therapy if needed.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
Battery critically low; charge immediately
Battery Critical
Audio Pause
Reset
Battery Low
Audio Pause
Reset
On/Off
Содержание ACTIV.A.C.
Страница 1: ...User Manual For Patients and Clinicians ACTIV A C THERAPY SYSTEM Rx Only...
Страница 2: ......
Страница 4: ......
Страница 5: ...For Patient Use ACTIV A C THERAPY SYSTEM...
Страница 6: ......
Страница 44: ......
Страница 45: ...For Clinician Use Only Patients Refer to Previous Section of This Manual ACTIV A C THERAPY SYSTEM...
Страница 46: ......
Страница 82: ...CLINICIAN 76...
Страница 83: ...CLINICIAN 77...