Canister Not Engaged Alarm
Medium Priority Alarm - This alarm screen appears when the canister is not properly installed. It is
indicated by a repeating audible tone.
Press
Audio Pause
to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will stop during this alarm.
This alarm is fixed by reinstalling the canister.
1. Press the canister latch release to remove the canister
(see
Canister
section).
2. Inspect the canister and therapy unit for:
•
Debris between the canister and therapy unit.
•
Two silicone seals (see
Canister
section).
•
Two canister bumpers (see
Canister
section).
3. If seals or bumpers are missing or damaged, contact
KCI.
4. Reinstall the canister. Make sure that it is fully engaged
and latched. An audible click should be heard when the
canister is properly installed.
5. Press
Reset
to return to the Patient Mode
Home
screen.
6. Press
On / Off
to restart therapy.
7. If the alarm continues, install a new canister (see
Canister Changes
section).
8. If the alarm cannot be fixed, call KCI or your doctor or
nurse.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
WARNING:
Ensure canister is in place.
Press ‘?’ for more information.
Audio Pause
Reset
Canister Not Engaged
Audio Pause
Reset
On/Off
Содержание ACTIV.A.C.
Страница 1: ...User Manual For Patients and Clinicians ACTIV A C THERAPY SYSTEM Rx Only...
Страница 2: ......
Страница 4: ......
Страница 5: ...For Patient Use ACTIV A C THERAPY SYSTEM...
Страница 6: ......
Страница 44: ......
Страница 45: ...For Clinician Use Only Patients Refer to Previous Section of This Manual ACTIV A C THERAPY SYSTEM...
Страница 46: ......
Страница 82: ...CLINICIAN 76...
Страница 83: ...CLINICIAN 77...