Find and Fix the Leak
Most leaks occur:
•
Where the drape meets the skin.
•
Where the SENSAT.R.A.C.
TM
Pad is attached to the drape.
•
At tubing connectors.
•
When the canister is not securely connected to the therapy unit.
To fix the leak:
1. Check the tubing connectors between the dressing and the canister. Make
sure they are properly locked.
2. Make sure the canister is properly installed onto the therapy unit. When the
canister is properly installed, it cannot be removed by gently pulling it away
from the unit.
3. Use the SEAL CHECK™ Leak Detector to find the leak at the dressing.
•
Make sure therapy is on.
•
Using light pressure, move your hand and fingers slowly around the
edges of the drape and SENSAT.R.A.C.
TM
Pad.
•
Watch the bar graph and listen to the audible tone. The bar will get
shorter and the frequency of the audible tone will decrease when the
leak is found.
4. When the leak is fixed, press
Exit
to return to the Patient Mode
Home
screen.
5. If the leak alarm persists or you cannot find the leak, call your doctor, nurse or
KCI.
Exit
Содержание ACTIV.A.C.
Страница 1: ...User Manual For Patients and Clinicians ACTIV A C THERAPY SYSTEM Rx Only...
Страница 2: ......
Страница 4: ......
Страница 5: ...For Patient Use ACTIV A C THERAPY SYSTEM...
Страница 6: ......
Страница 44: ......
Страница 45: ...For Clinician Use Only Patients Refer to Previous Section of This Manual ACTIV A C THERAPY SYSTEM...
Страница 46: ......
Страница 82: ...CLINICIAN 76...
Страница 83: ...CLINICIAN 77...