VII. CELL LIFE 

The Min iBum p generating source is consumed during normal 
operation. All sources are rated for up to 1000 test maximum, 250 
minimum. Once a source has been depleted, it may be replaced with a 
new source to provide another 1000 tests. MiniCal sources last for 5 
hours of use. Replacement sources may be stored in their containers in a 
cool, humid area for several years under most conditions. 

 

Different types of generating sources can be used with each instrument. 
The microprocessor in the MiniBump reads the type of gas of the source 
from a memory chip embedded in the source body.  It then adjusts itself 
for each type of gas output. For a complete list of correct source part 
numbers, please refer to the parts list to follow. 

 

To replace the generating source in the instrument, first remove the 
instrument cap located on the top of the instrument.  The generating source 
pulls directly up out of the source chamber. 

Do not twist or rotate the 

generating source while removing it from the source chamber or 
damage may occur to either the source pins or the source chamber 
sockets. 

The replacement source is replaced in reverse fashion.  Care 

must be taken when replacing the source to correctly line up the electrode 
pins with the corresponding sockets.  Do not force the source into the 
source chamber if it is not properly aligned, otherwise the pins may 
become damaged rendering the source and/or instrument useless. 
 

NOTE: Never attach a sample draw instrument directly to the end of 
the MiniBump delivery cup with no 'Y' or 'T' connector in between 
the two instruments.   

 

 

Содержание MiniBump

Страница 1: ...lephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0550 00 Rel IV AdvancedCalibrationDesigns Inc InstructionManual CalibrationGasBumpTester Advan...

Страница 2: ...nstrument correctly can lead to improper testing of your detection system This instrument conforms to the protection requirements of the EC DIRECTIVE 2014 30 EU on Electromagnetic Compatibility EMC in...

Страница 3: ...2...

Страница 4: ...5 Electrochemical Generating Source 5 Alkaline A Batteries 5 Microprocessor Based Circuitry 6 II START UP MINIBUMP SOURCE 7 III NORMAL OPERATION MINIBUMP SOURCE 8 IV START UP MINICAL SOURCE 9 V NORMAL...

Страница 5: ...4...

Страница 6: ...lectrochemically generated gas The Release IV MiniBump instruments are designed to operate at 5 LPM liters per minute Electrochemical Generating Source The electrochemical generating source contains a...

Страница 7: ...wered up the microprocessor checks for what type of generating source is inserted the gas type and concentration as well as if the source is a bump type source and the remaining source life It also mo...

Страница 8: ...ll begin running With a bump source installed the unit will run for an additional 20 seconds providing a bump of output gas to drive the connected sensor into alarm after which both red and green ligh...

Страница 9: ...rce is consumed If either condition occurs a RED LED will light indicating a non normal condition refer to the Non Normal Operation section that follows The calibration gas is delivered through a smal...

Страница 10: ...ting the unit is ready to calibrate With a MiniCal source installed the unit will continue to run in this state generating the concentration and type of gas specified by the generating source installe...

Страница 11: ...as is delivered through a small flexible hose attached to the source manifold cap The hoseis placed over around or in front of the gas sensor Alternatively the detector can be placed directly in front...

Страница 12: ...oved the RED LED will light continuously If the depleted source condition occurs during power up the RED LED will remain on for approximately 30 seconds and then the instrument will power down If the...

Страница 13: ...t numbers please refer to the parts list to follow To replace the generating source in the instrument first remove the instrument cap located on the top of the instrument The generating source pulls d...

Страница 14: ...altitude and temperature can cause changes in air density thus affecting the output of the MiniCal source The output was calibrated to STP standard temperature and pressure of sea level and 20 C To co...

Страница 15: ...he MiniBump instrument contain corrosive chemicals While it is not expected that the chemicals in the source or alkaline batteries will leak during normal operation it is recommended that both the sou...

Страница 16: ...e 510 0550 02 Hydrogen Sulfide MiniBump source 510 0570 02 Hydrogen Cyanide MiniBump source 510 0590 02 Hydrogen MiniBump source 510 0401 0x Chlorine MiniCal source specify PPM 1 25 510 0405 0x Hydrog...

Страница 17: ...d Calibration Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned toAdvanced Calibration Designs Inc at its manufa...

Страница 18: ...le electrochemical MiniCal sources and their standard warranty when installed in equipment manufactured and supplied by Advanced Calibration Designs Inc Chlorine Hydrogen Hydrogen Sulfide Hydrogen Cya...

Страница 19: ...anufactured by Advanced Calibration Designs Inc 2024 W McMillan Street Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Website www GoACD...

Страница 20: ...lephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0550 00 Rel IV AdvancedCalibrationDesigns Inc InstructionManual CalibrationGasBumpTester Advan...

Отзывы: