Section 4. Basic Operation

The manual covers all operations in depth in later sections for users who already
understand the basic operation. This section gives a brief and simple walk through of the
unit set up and general operation so that the user can quickly get started and accomplish
the primary test functions. For clarity, many additional features and options are not
addressed in this section of the manual. 

NOTE: The unit may automatically zero itself at 1 minute intervals so that any drift is
removed. There will be a slight audible click heard when this occurs.

Connecting To the Blood Pressure Monitor 

1. Select the appropriate adapter hose using the following table and the Contents
Checklist Illustration on previous page, which illustrates the hoses and can be used as a
guideline for selecting the appropriate adapter hose: 

Blood Pressure Monitor 

Adapter Hose 

DINAMAP Dinamap 
CAS Medical 

Luer 

Datascope Luer 
Philips Medical 

Philips Medical 

GE Medical GE 

Medical 

Invivo Luer 
Colin Medical 

Luer 

Criticare Luer 
SpaceLabs 

Luer or Universal 

All Other NIBP Monitors 

Universal 

If none of the above adapter hoses appear to match your tubing/cuff interface and if your
cuff has a standard “Universal-style” 5/32" ID rubber hose, you can use the universal
adapter hose. 

If you have questions regarding selection of the adapter hose, please contact: 

Technical Support Clinical Dynamics of CT, L.L.C. 
10 Capital Drive 

Wallingford, CT 06492

Phone: (203) 269-0090 

Fax: (203) 269-3402

Email: [email protected] 

2. Assuming you are using an adult cuff, wrap the cuff around the AccuPulse case just
as you would wrap it around a patient’s arm. You should be able to fit a pencil between
cuff and mandrel. Please see “Wrapping the Cuff Around the Cuff Support” illustration
below:

3. Connect the longest section of the adapter “T”
hose to the pressure port on the rear panel of the
AccuPulse.
4. Disconnect the cuff from the blood pressure
monitor hose. (As if you were going to replace the
cuff.) In the case of a dual-lumen hose, only one of
the hoses will be connected to the adapter.
5. Connect the cuff to one short side of the adapter
hose “T” and the hose to the other side of the “T”.
Ensure that an airtight seal is achieved on all three
couplings. Please refer to “Connecting AccuPulse to
the NIBP Monitor” illustration below:

BP Test Mode 

1. Please check all connections to assure a tight and leak-free seal. 
(NIBP Keypad ) BPTest mode is the startup default. If needed press the BPTest key to
change to this mode.
(Generic Keypad ) From the Main Menu, select NIBP Module <ENTER>, then BPTest
<ENTER> 

2. Initiate an NIBP reading on the monitor under test. 

Note:

Please take at least 3 readings at each setting or follow your test procedure. The

accuracy of the first reading after the cuff is wrapped is often less accurate by 1-2 mmHg.
This can be due to the cuff “seating” during the test. After your testing is complete,
compare the readings of the monitor with the BP Preset. 

Presets

To select a different Preset, Press the MENU key. 

Use the Left/Right Arrow keys to choose the appropriate tab. The presets are in 2 ranges

on tabs “BP-1” and “BP-2”.

Pressing an F1-F4 key selects that preset for use.

To leave press MENU again, or simply begin another reading on the monitor. With the
generic keypad the ESCAPE key is a third way to exit this menu.

NOTE: The last used preset will be remembered as the starting point and is stored across
power cycles.

Pulse Rate (PR) 

To select a different Pulse Rate (PR), Press the MENU key. 

Use the Left/Right Arrow keys to choose the appropriate tab. 

Pressing an F1-F3 key selects a preset value.

Pencil

Observe which side goes to patient 

Wrap excess cuff around
bottom of Cuff Support

To

BP

Monitor

To

BP

Cuff

Adapter

Hose

5

Содержание NIBP Simulator

Страница 1: ...ch the problem occurred Ideally a written problem description would be provided allowing for more efficient handling of your initial service request and the subsequent diagnosis and remedy of the problem 3 Your institution s complete name and address with a contact name and phone number 4 A purchase order number if the product needs non warranty service or if you are ordering spare parts Returning...

Страница 2: ...s It is imperative that the user read the AccuPulse Operation Manual carefully before operating AccuPulse NIBP Simulator Magnetic Media Caution A strong magnetic field exists at the left rear of the AccuPulse chassis Please avoid placing magnetically sensitive materials such as computer media in this vicinity Proper Battery Maintenance To maintain battery life when the unit shuts down due to low b...

Страница 3: ...uracy digital manometer Over Pressure Test Mode Automatically tests operation of the blood pressure monitor s pressure relief valve and records the set point when pressure is released by the relief valve Meter Mode In addition to the 0 1 mmHg resolution manometer this mode offers automated inflation to preset values or a user defined set pressure to verify static calibration across the entire rang...

Страница 4: ...to the current reading displays Min Max and a moving Average Set Meter Units mmHg kPa cmH20 psi The built in pump supports automated inflation in steps of 50mmHg The built in pump supports automated inflation to user defined set point adjustable on the fly When selecting another mode results displayed on various screens are maintained so that a user may run a test in one mode and come back to a pr...

Страница 5: ...stration below 3 Connect the longest section of the adapter T hose to the pressure port on the rear panel of the AccuPulse 4 Disconnect the cuff from the blood pressure monitor hose As if you were going to replace the cuff In the case of a dual lumen hose only one of the hoses will be connected to the adapter 5 Connect the cuff to one short side of the adapter hose T and the hose to the other side...

Страница 6: ...ation Engine operates in this mode delivering pressure pulses to the cuff The servo action begins when pressure in the system is 10mmHg If the user leaves the menu tabs displayed the servo will still start up if the system pressure becomes 10mmHg When it starts the software will change the screen to display the test being performed The figure below shows a typical BP Test Mode Screen with a maximu...

Страница 7: ...tal and Adult Modes Pressing the remaining three F Keys chooses one of three BP Presets for this patient type Figure BP 2 shows the first BPTest tab Because there are more than 3 presets there is a second BP tab Figure BP 3 shows the second BPTest tab The remaining presets may be selected using the second BPTest tab by pressing keys F1 through F4 PR Adjust Tab Figure PR shows the Pulse Rate adjust...

Страница 8: ...ould be consideration of the implications if the simulator is shared amongst multiple users or if a variety of equipment from different manufacturers needs to be tested with the same AccuPulse Plus unit This Trim function is provided for customers who want to dial in the AccuPulse for use with specific equipment The baseline generic oscillometric envelope is not based on any one monitor It is norm...

Страница 9: ... CalTables software option you can select a set of presets that have already been tuned by Clinical Dynamics to match the indicated Make Model of equipment This saves the user from always having to determine what the normal expected results should be for a given NIBP instrument With an appropriate CalTable selected the result typically agrees with the preset within just a few mmHg While in the BP ...

Страница 10: ... on using the Trim menu The MAP value displayed is a label only In the Trim Menu a user can adjust the value displayed to match the expected monitor results but can not directly move the underlying MAP point When making adjustments a minimum pulse pressure Sys Dia of 10 is enforced Attempting to move the systolic value below this spread will push the diastolic down as well Raising diastolic operat...

Страница 11: ...e upper right corner to the pressure on the NIBP monitor F1 Zero will zero the AccuPulse F2 Reset will reset the Leak Test NOTE Results remain displayed until a new test is started F3 Start start the Leak Test F4 Auto F4 Manual toggles between automatic using the internal pump or manual mode using the inflation bulb The button label states the current mode Auto is the default The Menu key opens th...

Страница 12: ... operations This provides the user with flexibility to choose the test conditions that suit their needs F1 F3 Set1 through Set3 select the settings to use or configure Left Right arrow keys select the parameter to adjust Up Down arrow keys adjust the value of the highlighted parameter Target Pressure range is 50 300 mmHg Test Time range is 20 120 seconds Changes are saved immediately In manual mod...

Страница 13: ... stabilize is strongly recommended F1 Zero will zero the manometer F2 Reset will reset the Max Min and Avg F3 Cal Enters the Calibration check mode This provides incrementing steps of 50 mmHg F4 SetP Enters the Set Pressure mode where the user can interactively set a target inflation value The user may Zero the pressure sensor with the unit open to the atmosphere to remove any drift The unit also ...

Страница 14: ...o be made while the test is active Press Start to begin the test When the target pressure is reached Accupulse will try to maintain the pressure at this value The corrections made by AccuPulse may cause a slight change just when you don t want it Pressing Disable will turn off adjustments allowing the pressure to naturally settle and drift down smoothly If the pressure should drop too low pressing...

Страница 15: ...t begins Beyond the timeout a hardware limit will force the unit off when battery is below 11V The battery voltage discharge curve is relatively slow and smooth until it approaches the end of its useful charge Then the voltage will fall rapidly The Battery is a LiOn pack and an intelligent charge circuit is used The battery will be charged as needed and as fast as possible within safety bounds It ...

Страница 16: ...mmands That is typically commands and command feedback are accompanied by similar syntax This section explains the commands in detail A less detailed short form listing of the most commonly used commands is presented at the end which can be used for fast reference Set up protocol Before developing a software communications driver for use in your AccuPulse application it is recommended that the dev...

Страница 17: ...ecommended These commands turn processing of the physical keypad off on so that the user does not have to worry about keypad inputs interfering with the execution of a remote control session Help This command returns a listing of available commands in the current software These may vary based upon the configuration options loaded in the unit Send HELP Recv Available Remote Commands TEST_MODE SET_S...

Страница 18: ...mber 46 and is provided in all NIBP units Numbers 1 45 can be populated with CalTables created by Clinical Dynamics to match specific equipment Currently the first 10 are used Send SET_CALTABLE CR 1 46 CR Recv success out of range Get CalTable Send GET_CALTABLE CR Recv xx Note CalTable index is valid only in BP Mode Get BP Parameters Send GET_BP_PARAMS CR Recv PRESET_TYPE BP BP or UTP PATIENT_TYPE...

Страница 19: ...at a group of them may be controlled remotely Software configured to control a group of devices can requisition each device for its informal name or a more formal ID such as Factory1 for instance This way it does not matter which USB port is used to communicate with a device even if it is on a Radio linked USB hub it just matters that we know which device ID we are talking to Some users will prefe...

Страница 20: ...it answers with it s current settings SET_PRESET 1 1 1 1 is ADULT preset 1 typically 60 30 ADULT presets are 1 7 SET_PRESET 0 6 0 6 is NEO preset 6 typically 150 120 NEO presets are 1 6 SET_PR 123 Pulse Rate 123 Range 15 330 SET_AMP 99 Amplitude 99 Range 0 150 most users should leave this at default of 100 SET_SHFT 30 shift preset by Negative 30 mmHg SET_SHFT 50 shift preset by Positive 50 mmHg Sh...

Отзывы: