Commonly Used Remote Control Commands Short Listing
<>
(Initiates the communication, wake simulator comm port)
<KEYPAD_OFF>
(locks out front panel keypad use)
<KEYPAD_ON>
(enables it again)
General Buttons
These button commands operate throughout all modes.
Variations for alternate keypads
Generic Keypad
Bench Top Unit
<BUTTON>
BP
ESC
<BUTTON>
LEAK
SCROLL_UP PREV
<BUTTON>
OVERP
SCROLL_DOWN NEXT
<BUTTON>
METER
ENTER
<BUTTON>
MENU
<BUTTON>
F1
<BUTTON>
F2
<BUTTON>
F3
<BUTTON>
F4
<BUTTON>
UP
<BUTTON>
DWN
<BUTTON>
LEFT
<BUTTON>
RIGHT
There are 2 direct commands to read the manometer
<MAN_STR>
( Continuously stream readings )
[ON/OFF]
<MAN_SMP>
( Single reading sample )
The following bypass some button sequences and set operating parameters
with direct commands. These should produce cleaner final coded command
sequences and reduce effort for the most often changed settings.
BP Test Settings
<GET_BP_PARAMS>
(unit answers with it's current settings)
<SET_PRESET>
1,1
( 1,1 is ADULT, preset#1 = typically 60/30 )
(ADULT presets are 1-7)
<SET_PRESET>
0,6
( 0,6 is NEO, preset#6 = typically 150/120 )
(NEO presets are 1-6)
<SET_PR>
123
( Pulse Rate = 123 Range 15-330 )
<SET_AMP>
99
( Amplitude = 99
Range 0-150% )
( most users should leave this at default of 100%)
<SET_SHFT>
-30
( shift preset by Negative 30 mmHg )
<SET_SHFT>
50
( shift preset by Positive 50 mmHg )
( Shift range -50 to 100 mmHg, OR, instrument
Max/Min limits)
The following are less often changed operating parameters. Some are only
available if the option is loaded.
BP Test Configuration Settings
<SET_BP_MODE>
*OPTION*
BP_PRESET
(or UTP_MODE if optional User Defined mode was
loaded)
<SET_CALTABLE>
*OPTION*
5
( digit 1-45 , only populated slots can be selected )
<SET_CALTABLE>
46
( 46 = Generic, it is the default included in all units
)
<SET_UTP>
*OPTION*
32
( digit 1-45 )
Special Commands not typically needed
<GET_SERIAL_NUM>
(returns the unit Serial Number)
<I>
(returns the Software version)
<GET_DEVICE_ID>
(returns USB/FTDI ID if set, or if not set “* “ )
<SET_DEVICE_ID>
CDCT #1
(sets USB/FTDI ID to " CDCT #1 ")
20