AccuPulse NIBP Simulator Скачать руководство пользователя страница 15

Power Configuration

This sets inactivity self turn off and other power saving options. The speaker may also be
disabled here.

Regardless of any settings made here, the unit will always shutdown in 10 minutes when
the Low Battery alert is flashing. Connecting the AC/DC charge adapter during this time
will prevent shutdown.

F1 (Toggle Sound) Turns the speaker Off/On.

F2 (Disp LED) toggles the display LED back light between always on, and inactivity timer

F3 (Idle Timer) toggles the unit idle shutdown timer On/Off

Idle Timer Shutdown will enable the unit to shut itself off after 30 minutes of inactivity.
By default this timer is turned off.

The display LED timer will turn off the backlight LED after 10 minutes of inactivity.
Pressing any key will reset the timer and turn the light back on. By default this timer is
turned off.

Both timers are reset with any button press.

Battery / Charger Voltage Status 

This will indicate the current power source in use, Battery or Charger, and its voltage.

When the Battery voltage falls to 13.95V a Low Battery warning will flash on every screen.
Once it starts flashing the unit will shutdown in 10 minutes, allowing enough time for
the user to complete the current task or plug in the external power charger.

The low battery alert is a blinking reverse-video bar that is drawn on the screen.

To fully charge the internal battery takes about 3.5 hours.
The nominal battery voltage displayed at full charge is approximately 16.5v
Low Battery alert will come on when the battery drops to 13.95V.
Shutoff will occur 10 minutes after the Low Battery alert begins.
Beyond the timeout, a hardware limit will force the unit off when battery is below 11V.

The battery voltage discharge curve is relatively slow and smooth until it approaches the

end of its useful charge. Then the voltage will fall rapidly.

The Battery is a LiOn pack and an intelligent charge circuit is used. The battery will be
charged as needed and as fast as possible within safety bounds. It will charge anytime
the charge adapter is being used, whether the simulator is turned on or not. While the
simulator is in use charging may be slightly slower causing some increase in the time
needed to fully recharge the battery.

Section 6. Specifications 

AccuPulse Plus NIBP Simulator Technical Specifications 

AccuPulse NIBP CalTables™ 

Datascope Accutorr Plus
GE Medical Dash 4000
GE Medical Dinamap Pro
Philips Medical M3
Welch-Allyn Propaq Encore
Philips Medical VS1
Colin Medical
Criticare
CAS Medical
SunTech

Please contact the factory for the availability of other calibration tables, or visit our
web site for updates at www.clinicaldynamics.com 

BP Presets
Systolic/Diastolic (MAP):

Adult:

Neonatal:

240/190

150/120

200/150

120/90 

150/100 

100/70

120/80

80/50

100/70

60/30

80/50

35/15

60/30

Exact Diastolic, Systolic and MAP values are a function of the specific CalTables™.
The BP Presets above are taken from the GENERIC CalTable™. 

Accuracy: ± 0.5 mmHg 

Oscillometric Pulse

Pulse Amplitude Range: 0-150% of nominal
Pulse Amplitude Resolution: 1%

Pulse Rate

Pulse Rate Range: 15-330 bpm
Pulse Rate Accuracy: ±0.5 bpm, 15-300 bpm, ±1 bpm 301-330 bpm

Displayed/Measured Parameters 

Dynamic NIBP Cuff Pressure Waveform

Digital Manometer

Pressure Range:

0.0 to 400.0 mmHg

Accuracy:

0 – 400 mmHg ± 0.5 mmHg

Resolution:

±0.1 mmHg 

Power Configuration screen

Fig. P1

Low Battery Alert sample screen

Fig. LB1

15

Содержание NIBP Simulator

Страница 1: ...ch the problem occurred Ideally a written problem description would be provided allowing for more efficient handling of your initial service request and the subsequent diagnosis and remedy of the problem 3 Your institution s complete name and address with a contact name and phone number 4 A purchase order number if the product needs non warranty service or if you are ordering spare parts Returning...

Страница 2: ...s It is imperative that the user read the AccuPulse Operation Manual carefully before operating AccuPulse NIBP Simulator Magnetic Media Caution A strong magnetic field exists at the left rear of the AccuPulse chassis Please avoid placing magnetically sensitive materials such as computer media in this vicinity Proper Battery Maintenance To maintain battery life when the unit shuts down due to low b...

Страница 3: ...uracy digital manometer Over Pressure Test Mode Automatically tests operation of the blood pressure monitor s pressure relief valve and records the set point when pressure is released by the relief valve Meter Mode In addition to the 0 1 mmHg resolution manometer this mode offers automated inflation to preset values or a user defined set pressure to verify static calibration across the entire rang...

Страница 4: ...to the current reading displays Min Max and a moving Average Set Meter Units mmHg kPa cmH20 psi The built in pump supports automated inflation in steps of 50mmHg The built in pump supports automated inflation to user defined set point adjustable on the fly When selecting another mode results displayed on various screens are maintained so that a user may run a test in one mode and come back to a pr...

Страница 5: ...stration below 3 Connect the longest section of the adapter T hose to the pressure port on the rear panel of the AccuPulse 4 Disconnect the cuff from the blood pressure monitor hose As if you were going to replace the cuff In the case of a dual lumen hose only one of the hoses will be connected to the adapter 5 Connect the cuff to one short side of the adapter hose T and the hose to the other side...

Страница 6: ...ation Engine operates in this mode delivering pressure pulses to the cuff The servo action begins when pressure in the system is 10mmHg If the user leaves the menu tabs displayed the servo will still start up if the system pressure becomes 10mmHg When it starts the software will change the screen to display the test being performed The figure below shows a typical BP Test Mode Screen with a maximu...

Страница 7: ...tal and Adult Modes Pressing the remaining three F Keys chooses one of three BP Presets for this patient type Figure BP 2 shows the first BPTest tab Because there are more than 3 presets there is a second BP tab Figure BP 3 shows the second BPTest tab The remaining presets may be selected using the second BPTest tab by pressing keys F1 through F4 PR Adjust Tab Figure PR shows the Pulse Rate adjust...

Страница 8: ...ould be consideration of the implications if the simulator is shared amongst multiple users or if a variety of equipment from different manufacturers needs to be tested with the same AccuPulse Plus unit This Trim function is provided for customers who want to dial in the AccuPulse for use with specific equipment The baseline generic oscillometric envelope is not based on any one monitor It is norm...

Страница 9: ... CalTables software option you can select a set of presets that have already been tuned by Clinical Dynamics to match the indicated Make Model of equipment This saves the user from always having to determine what the normal expected results should be for a given NIBP instrument With an appropriate CalTable selected the result typically agrees with the preset within just a few mmHg While in the BP ...

Страница 10: ... on using the Trim menu The MAP value displayed is a label only In the Trim Menu a user can adjust the value displayed to match the expected monitor results but can not directly move the underlying MAP point When making adjustments a minimum pulse pressure Sys Dia of 10 is enforced Attempting to move the systolic value below this spread will push the diastolic down as well Raising diastolic operat...

Страница 11: ...e upper right corner to the pressure on the NIBP monitor F1 Zero will zero the AccuPulse F2 Reset will reset the Leak Test NOTE Results remain displayed until a new test is started F3 Start start the Leak Test F4 Auto F4 Manual toggles between automatic using the internal pump or manual mode using the inflation bulb The button label states the current mode Auto is the default The Menu key opens th...

Страница 12: ... operations This provides the user with flexibility to choose the test conditions that suit their needs F1 F3 Set1 through Set3 select the settings to use or configure Left Right arrow keys select the parameter to adjust Up Down arrow keys adjust the value of the highlighted parameter Target Pressure range is 50 300 mmHg Test Time range is 20 120 seconds Changes are saved immediately In manual mod...

Страница 13: ... stabilize is strongly recommended F1 Zero will zero the manometer F2 Reset will reset the Max Min and Avg F3 Cal Enters the Calibration check mode This provides incrementing steps of 50 mmHg F4 SetP Enters the Set Pressure mode where the user can interactively set a target inflation value The user may Zero the pressure sensor with the unit open to the atmosphere to remove any drift The unit also ...

Страница 14: ...o be made while the test is active Press Start to begin the test When the target pressure is reached Accupulse will try to maintain the pressure at this value The corrections made by AccuPulse may cause a slight change just when you don t want it Pressing Disable will turn off adjustments allowing the pressure to naturally settle and drift down smoothly If the pressure should drop too low pressing...

Страница 15: ...t begins Beyond the timeout a hardware limit will force the unit off when battery is below 11V The battery voltage discharge curve is relatively slow and smooth until it approaches the end of its useful charge Then the voltage will fall rapidly The Battery is a LiOn pack and an intelligent charge circuit is used The battery will be charged as needed and as fast as possible within safety bounds It ...

Страница 16: ...mmands That is typically commands and command feedback are accompanied by similar syntax This section explains the commands in detail A less detailed short form listing of the most commonly used commands is presented at the end which can be used for fast reference Set up protocol Before developing a software communications driver for use in your AccuPulse application it is recommended that the dev...

Страница 17: ...ecommended These commands turn processing of the physical keypad off on so that the user does not have to worry about keypad inputs interfering with the execution of a remote control session Help This command returns a listing of available commands in the current software These may vary based upon the configuration options loaded in the unit Send HELP Recv Available Remote Commands TEST_MODE SET_S...

Страница 18: ...mber 46 and is provided in all NIBP units Numbers 1 45 can be populated with CalTables created by Clinical Dynamics to match specific equipment Currently the first 10 are used Send SET_CALTABLE CR 1 46 CR Recv success out of range Get CalTable Send GET_CALTABLE CR Recv xx Note CalTable index is valid only in BP Mode Get BP Parameters Send GET_BP_PARAMS CR Recv PRESET_TYPE BP BP or UTP PATIENT_TYPE...

Страница 19: ...at a group of them may be controlled remotely Software configured to control a group of devices can requisition each device for its informal name or a more formal ID such as Factory1 for instance This way it does not matter which USB port is used to communicate with a device even if it is on a Radio linked USB hub it just matters that we know which device ID we are talking to Some users will prefe...

Страница 20: ...it answers with it s current settings SET_PRESET 1 1 1 1 is ADULT preset 1 typically 60 30 ADULT presets are 1 7 SET_PRESET 0 6 0 6 is NEO preset 6 typically 150 120 NEO presets are 1 6 SET_PR 123 Pulse Rate 123 Range 15 330 SET_AMP 99 Amplitude 99 Range 0 150 most users should leave this at default of 100 SET_SHFT 30 shift preset by Negative 30 mmHg SET_SHFT 50 shift preset by Positive 50 mmHg Sh...