5

Basic Weighing Function

Features

– Zeroing the balance

You can zero the balance/scale within the entire weighing
range, up to the maximum capacity.

Preparation

§

Switch on the balance/scale: press the (ON/OFF) key

§

If necessary, zero the balance/scale: press the (ZERO) key

$

If necessary, change the configuration settings: 
see the chapter entitled “Configuration”

$

If desired, load the factory settings: 
see the chapter entitled “Configuration,” menu item 9.– 1

Additional functions:

$

Switching off the balance/scale: press the (ON/OFF) key

$

Battery operation: automatic shut-off after 2,5 or 10 min-
utes; see chapter on configuration. Example: 2 minutes.
If the weight readout remains unchanged and no keys are
pressed for at least two minutes, the battery symbol

“ starts flashing. After another 2 seconds, the 

balance shuts off automatically, unless a key is pressed.

Description of the Keys

(ON/OFF)

On/off key: switches the balance/scale on and
off or switches it to the standby mode.
Battery operation: on; turns backlight on; off

(ZERO)

Zeros the balance; press and hold 2 seconds:
opens the application menu

(CAL)

Starts calibration/adjustment

(F)

Starts an application program; 
Scrolling in application menu, configration
menu and calibration menu

(ENTER)

Confirms the selected setting;
Exits application, configuration & calibration
menu if key is pressed and held for more than 
2 seconds.

(PRINT)

Generates a printout or data output

Operation

Example:

Determine weight of sample

Step

Key (or instruction)

Display

1. Switch on the balance/scale

(ON/OFF)

Self-test runs
Display: Software version

r 31.01 

2. Open the in-use cover and leave open

while weighing

3. Place container on the balance/scale

52.0

g

(in this example, 52 g)

4

. Zero the balance/scale

(ZERO)

0.0

g

5. Place sample in container on balance/scale

150.2

g

(in this example, 150,2 g).

Visit www.balances.com for the best prices on Acculab scales

www.balances.com is the original deep discount distributor

Содержание 98648-013-61

Страница 1: ...ULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elettroniche di precisione Balanzas electr nicas de precisi n Vis...

Страница 2: ...autions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance scale Do not use this equipment in hazardous areas Make sure...

Страница 3: ...top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct sunlight Drafts from open windows and doors Extreme vibrati...

Страница 4: ...ve the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cover into position until it snaps into place Do not throw aw...

Страница 5: ...e Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on off ZERO Zeros the balance press and hold 2 seconds opens the...

Страница 6: ...version factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pounds 0 00220462260 lb 6 ounce Ounces 0 03527396200 oz 7 tryo...

Страница 7: ...uantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 10 10 20 30 100 F 2 sec 8 Confirm selected reference sample q...

Страница 8: ...6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 factory setting F repeatedly 8 Confirm selected no of decimal p...

Страница 9: ...alance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is locked 7 Unload the balance scale the weight value remains d...

Страница 10: ...ce scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is displayed symbol flashes Note symbol remains on indicating stored v...

Страница 11: ...display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sample from balance scale 10 Determine weight of sample in liq...

Страница 12: ...libration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on the display 6 Place the required calibration weight on the...

Страница 13: ...1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibration adjustment Function of the CAL key 1 5 2 Linearization...

Страница 14: ...aded Unload the balance scale appErr Weight is too light or Increase the weight on the balance scale there is no sample on the balance scale with application in use PrtErr Data interface for Contact t...

Страница 15: ...eable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100 Readability g 0 1 0 1 0 1 0 1 0...

Страница 16: ...Standards Limitation of emissions In accordance with product standard EN 61326 1 Class B residential area Defined immunity to interference in accordance with product standard EN 61326 1 minimum test r...

Отзывы: