Error codes are shown on the main display for approx. 2 seconds. The program then returns automatically to the previous mode.

Display/Problem

Cause

Solution

No segments appear

No power available

Check the power supply

on the display 

The AC adapter is not plugged in

Plug in the AC adapter

Battery is drained

Replace battery; recharge battery using 
external charger

oL

The load exceeds the balance/scale 

Unload the balance/scale

capacity

UL

Weighing pan not in place

Place the weighing pan
on the balance/scale

Something is touching

Move the object that is 

the weighing pan

touching the weighing pan

disErr

Display overflow: 

Reduce load on the balance/scale

Value cannot be shown 
on the display 

calErr

Calibration parameter not met; e.g.:

Calibrate only when zero is displayed

– Balance/scale not zeroed

Press (ZERO) to tare the balance/scale

– Balance/scale is loaded

Unload the balance/scale

appErr

Weight is too light or 

Increase the weight on the balance/scale

there is no sample on the balance/scale
with application in use

PrtErr

Data interface for 

Contact the Acculab

printing is blocked

customer service center 

bal.Err

Balance/scale loaded or defecteive

Unload balance/scale before switching on

when power was turned on

or contact Acculab custumer service

sys.Err

Balance/scale defective 

Contact Acculab custumer service 

Max. weighing capacity is less 

The balance/scale was switched on

Place the weighing pan on the 

than indicated under “Specifications”

without the weighing pan in place

balance/scale and press (ON/OFF)

The weight readout is obviously wrong

The balance/scale was not calibrated/

Calibrate/adjust the balance/scale

adjusted before weighing
Balance/scale not zeroed

Zero the balance/scale 

If any other errors occur, contact your local Acculab customer service center.

Error Codes

Visit www.balances.com for the best prices on Acculab scales

www.balances.com is the original deep discount distributor

Содержание 98648-013-61

Страница 1: ...ULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elettroniche di precisione Balanzas electr nicas de precisi n Vis...

Страница 2: ...autions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance scale Do not use this equipment in hazardous areas Make sure...

Страница 3: ...top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct sunlight Drafts from open windows and doors Extreme vibrati...

Страница 4: ...ve the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cover into position until it snaps into place Do not throw aw...

Страница 5: ...e Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on off ZERO Zeros the balance press and hold 2 seconds opens the...

Страница 6: ...version factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pounds 0 00220462260 lb 6 ounce Ounces 0 03527396200 oz 7 tryo...

Страница 7: ...uantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 10 10 20 30 100 F 2 sec 8 Confirm selected reference sample q...

Страница 8: ...6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 factory setting F repeatedly 8 Confirm selected no of decimal p...

Страница 9: ...alance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is locked 7 Unload the balance scale the weight value remains d...

Страница 10: ...ce scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is displayed symbol flashes Note symbol remains on indicating stored v...

Страница 11: ...display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sample from balance scale 10 Determine weight of sample in liq...

Страница 12: ...libration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on the display 6 Place the required calibration weight on the...

Страница 13: ...1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibration adjustment Function of the CAL key 1 5 2 Linearization...

Страница 14: ...aded Unload the balance scale appErr Weight is too light or Increase the weight on the balance scale there is no sample on the balance scale with application in use PrtErr Data interface for Contact t...

Страница 15: ...eable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100 Readability g 0 1 0 1 0 1 0 1 0...

Страница 16: ...Standards Limitation of emissions In accordance with product standard EN 61326 1 Class B residential area Defined immunity to interference in accordance with product standard EN 61326 1 minimum test r...

Отзывы: