background image

13

To configure the balance/scale; i.e., adapt the balance/scale to individual requirements.

Step

Key (or instruction)

Display

1. Switch off the balance/scale 

(ON/OFF)

2. Switch on the balance/scale and 

(ON/OFF)

while all segments are displayed

(ZERO) briefly

Navigation in the Setup Menu

Key

Press briefly

Press and hold

(ENTER)

Menu level: move to the right (cyclical)

Confirm setting

(F)

Menu item: Scrolling 

(ZERO)

Menu level: Move to the left

Save settings and exit Setup

Parameter Settings (Overview)

ο

Factory setting

√ 

User-defined setting

Setup

1

Weighing

1.1. Adapt filter

1.1.1

Very stable conditions

menu

1.1.2

ο

Stable conditions

1.1.3

Unstable conditions

1.1.4

Very unstable conditions

1.2. Application filter

1.2.1

ο

Final readout

1.2.2

Filling

1.3. Stability range

1.3.1

1/4 digit

1.3.2

1/2 digit

1.3.3

1 digit

1.3.4

ο

2 digits

1.3.5

4 digits

1.5. Calib./adjust./linearization:

1.5.1

ο

Calibration/adjustment

Function of the (CAL) key

1.5.2

Linearization: for service personnel only

1.5.3

Key blocked

1.6. Auto zero

1.6.1

ο

On

1.6.2

Off

1.7. 1st weight unit, or 2nd unit in 

1.7.1

User-defined unit; see 

Toggle Weight Units app.

to 1.7.23

“Toggling between Weight Units”

5. and 6. 

Only relevant with built-in data interface: see corresponding interface description 

8. Additional

8.1. Block key functions

8.1.1

All keys blocked except for (ON/OFF)

functions

and (ZERO)

8.1.2

ο

All keys unblocked

8.2. Automatic shut-off 

8.2.1

ο

After 2 minutes

8.2.2

After 5 minutes

8.2.3

After 10 minutes

9. Reset menu

9.1. Factory settings

9.1.1

Restore

9.1.2

ο

Do not restore

Configuration (Setup Menu)

Visit www.balances.com for the best prices on Acculab scales

www.balances.com is the original deep discount distributor

Содержание 98648-013-61

Страница 1: ...ULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elettroniche di precisione Balanzas electr nicas de precisi n Vis...

Страница 2: ...autions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance scale Do not use this equipment in hazardous areas Make sure...

Страница 3: ...top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct sunlight Drafts from open windows and doors Extreme vibrati...

Страница 4: ...ve the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cover into position until it snaps into place Do not throw aw...

Страница 5: ...e Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on off ZERO Zeros the balance press and hold 2 seconds opens the...

Страница 6: ...version factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pounds 0 00220462260 lb 6 ounce Ounces 0 03527396200 oz 7 tryo...

Страница 7: ...uantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 10 10 20 30 100 F 2 sec 8 Confirm selected reference sample q...

Страница 8: ...6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 factory setting F repeatedly 8 Confirm selected no of decimal p...

Страница 9: ...alance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is locked 7 Unload the balance scale the weight value remains d...

Страница 10: ...ce scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is displayed symbol flashes Note symbol remains on indicating stored v...

Страница 11: ...display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sample from balance scale 10 Determine weight of sample in liq...

Страница 12: ...libration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on the display 6 Place the required calibration weight on the...

Страница 13: ...1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibration adjustment Function of the CAL key 1 5 2 Linearization...

Страница 14: ...aded Unload the balance scale appErr Weight is too light or Increase the weight on the balance scale there is no sample on the balance scale with application in use PrtErr Data interface for Contact t...

Страница 15: ...eable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100 Readability g 0 1 0 1 0 1 0 1 0...

Страница 16: ...Standards Limitation of emissions In accordance with product standard EN 61326 1 Class B residential area Defined immunity to interference in accordance with product standard EN 61326 1 minimum test r...

Отзывы: