Chapter 3, Preparing the devices
3-20
Accu-Chek Smart Pix user's manual
3.17
Note on time settings for the Accu-Chek insulin pumps
If you have to correct the time set in your Accu-Chek insulin
pump, please observe the following:
Do
not
change the time on your Accu-Chek insulin
pump if you have programmed time-based changes
of insulin delivery (e.g. temporary basal rate, extended
bolus or multiwave bolus) and this change is still in
effect (i.e. the programmed period of time has not yet
elapsed).
In this case, the Accu-Chek Smart Pix system may not be
able to analyse the stored insulin pump data. Therefore, only
change the time set in your insulin pump if it is in STOP
mode.
If you want to reset the time programmed in your Accu-Chek
insulin pump (e.g. at the end of the summer time or due to
time zone changes when travelling to Western directions)
and to analyse the existing data using the Accu-Chek
Smart Pix system:
Before
changing the time, read the existing data from
your Accu-Chek insulin pump. The next time data is
read, the Accu-Chek Smart Pix system only analyses
the insulin pump data which was stored
after
the
time was reset.
If you use your insulin pump together with a blood glucose
meter and want to analyse the data of both devices using the
Accu-Chek Smart Pix system,
always
change the time
for all
devices
at once. This ensures that the information
represented in the reports is assigned correctly. Combined
reports are only created for devices having the same date
and time settings.
i
i