Glucose Control Testing
58
Confirming or selecting the lot
number for control solutions
Once you have selected the level, you are asked to
con firm or enter the lot number of the control solution.
Com pare the number displayed by the meter to the
number on the label of the control solution bottle.
1
Select the lot number as follows:
n
To read the lot number from the control solution
bottle via barcode scanner, press and
release
. Follow the instructions for
scanning IDs (see page 35). *
n
To use the preselected number displayed by the
meter, touch
to confirm.
n
To use a different number than the lot number
displayed, touch
to open the keypad and
enter the number manually.
2
Touch
to confirm the selected control lot
number.
The meter can be configured so that manual
confirmation is not necessary. In this setup, only the lot
number is displayed. Additional options are not
available.
The meter can be configured so that lot numbers are
entered via barcode scanner only.
For additional information about storing lot numbers for
control solutions, see page 74.
Control Test
Control L2 (Hi)
123456?
Use Control Lot
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Control Lot
12:48 pm
09/18/13
Содержание Inform 2
Страница 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ......
Страница 11: ... 10 This page intentionally left blank ...
Страница 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Страница 91: ...Linearity Testing 90 This page intentionally left blank ...
Страница 99: ...Proficiency Testing 98 This page intentionally left blank ...
Страница 149: ...General Product Information 148 This page intentionally left blank ...
Страница 167: ...Appendix 166 This page intentionally left blank ...
Страница 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Страница 197: ... 196 This page intentionally left blank ...
Страница 198: ......