Initial Startup
114
On the second settings screen, you can select options for
handling lot numbers with glucose control and linearity
tests.
1
Touch the desired option to enable it:
n
Control Lot Verification
–
Display only
: The lot number is shown on-screen
but the operator cannot confirm it or select a
different lot number.
–
Confirmation
: The operator must confirm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
–
Lot entry
: The operator must enter the lot number.
–
Scan only
: Lot numbers can be verified via barcode
scanner only.
n
Linearity Lot Verification
–
Display only
: The lot number is shown on-screen
but the operator cannot confirm it or select a
different lot number.
–
Confirmation
: The operator must confirm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
–
Lot entry
: The operator must enter the lot number.
2
Touch
to save the settings (for both settings
screens), or
3
Touch
to return to the first settings screen.
Reagent Options
Control Lot Verifi cation
Confi rmation
Display only
Scan only
Linearity Lot Verifi cation
Display only
Confi rmation
Lot entry
Lot entry
12:48 pm
09/18/13
Содержание Inform 2
Страница 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ......
Страница 11: ... 10 This page intentionally left blank ...
Страница 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Страница 91: ...Linearity Testing 90 This page intentionally left blank ...
Страница 99: ...Proficiency Testing 98 This page intentionally left blank ...
Страница 149: ...General Product Information 148 This page intentionally left blank ...
Страница 167: ...Appendix 166 This page intentionally left blank ...
Страница 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Страница 197: ... 196 This page intentionally left blank ...
Страница 198: ......