Testing blood glucose
Farbe: P 021 – Prüfmittelnummer: 201
5.2.3 Notes
Only operate your Accu-Chek Go meter with Roche Diagnostics
GmbH-approved Accu-Chek Go test strips. If you use the wrong
test strips, your results will be incorrect. Incorrect results can
cause the wrong therapeutic decision to be taken and so lead to
serious adverse health effects.
The test strips are sensitive to humidity. Make sure your hands
are completely dry when you take the test strip from the con-
tainer. Close the test strip container again with the original cap
immediately after removing the test strip. The cap contains a dry-
ing agent to protect the test strips against moisture. If moisture
enters the container through handling with moist hands or if the
test strip container is left open, the drying agent ceases to act.
Test strips that are unusable may produce incorrect results.
Incorrect results can cause the wrong therapeutic decision to be
taken and so lead to serious adverse health effects.
Be sure not to bend the test strip when inserting it in the meter.
Bending the test strip can produce an incorrect result or cause an
E-1 (see page 172) or E-6 (see page 176) error message to be
displayed.
Содержание GO
Страница 1: ...BLOOD GLUCOSE METER Instructions for use ...
Страница 10: ...10 Introduction 1 Introduction 1 1 The Accu Chek Go blood glucose meter at a glance h g f e a b c d l k j g i ...
Страница 42: ...42 Entering settings Month Day ...
Страница 66: ...66 Entering settings Entering alarm times Hours Minutes Alarm 2 ...
Страница 187: ...187 Patents 18 18 Patents US 5 463 467 US 5 366 609 US 6 707 554 ...
Страница 204: ...204 ...
Страница 205: ......
Страница 207: ...Warranty Card Name Street City post code Telephone No Date Signature ...