138
Checking your meter
Farbe: P 021 – Prüfmittelnummer: 201
Have you observed the proper storage conditions for the
meter, test strips and control solution (see Chapter 12 “Mea-
surement and storage conditions”, and the package inserts
that came with the test strips and control solutions)?
yes
Are the test strip guide and measurement optics clean (see
Chapter 9 “Cleaning your meter”)?
yes
Are the control solution and test strips past their expiry date
(look next to the symbol on the bottle and test strip con-
tainer label and look to see if
exp
is shown in the display)?
no
Has the control solution been open for less than three
months?
Once opened, control solutions are only stable for three
months and must not be used after this period.
yes
If you have responded correctly to all these questions and still receive
implausible results or error messages, please call your local customer
support and service centre (see Chapter 19).
You may also receive implausible results or error messages if your
meter has been dropped. In this instance, too, please call your local cus-
tomer support and service centre.
Содержание GO
Страница 1: ...BLOOD GLUCOSE METER Instructions for use ...
Страница 10: ...10 Introduction 1 Introduction 1 1 The Accu Chek Go blood glucose meter at a glance h g f e a b c d l k j g i ...
Страница 42: ...42 Entering settings Month Day ...
Страница 66: ...66 Entering settings Entering alarm times Hours Minutes Alarm 2 ...
Страница 187: ...187 Patents 18 18 Patents US 5 463 467 US 5 366 609 US 6 707 554 ...
Страница 204: ...204 ...
Страница 205: ......
Страница 207: ...Warranty Card Name Street City post code Telephone No Date Signature ...