8.2 Troubleshooting the ACCU‑CHEK Aviva Combo Meter
Display Shows Possible Cause(s)
Possible Solution(s)
The display is
blank or the meter
will not turn on.
j
Batteries are dead.
j
Install new batteries. See chapter 7, “care and
Maintenance.”
j
Display is damaged.
j
contact Accu-chek Pump Support at
1-800-688-4578.
j
Meter is defective.
j
contact Accu-chek Pump Support at
1-800-688-4578.
j
extreme temperature – the
temperature is outside the
meter operating range.
j
Move the meter to an area with proper temperature.
Wait five minutes before turning on the meter. Do
not artificially heat or cool the meter.
Bluetooth
icon
flashing
j
communication ended due to
a button press on the pump.
j
check the pump and continue operation of the
pump manually.
j
ensure the pump has
Bluetooth
wireless technology
turned on.
j
Meter and pump are outside
of communication range.
j
ensure pump is within communication range.