240
Display Shows Possible Cause(s)
Possible Solution(s)
Bolus Delivery
unavailable
j
The meter cannot
communicate with the pump.
j
Retry or set to Manual Pump.
j
ensure the meter and pump have
Bluetooth
wireless
technology turned on.
j
ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j
The pump is currently in Stop
mode.
j
Take the pump out of Stop mode.
j
The pump is currently
delivering a bolus.
j
Allow the current bolus delivery to complete before
delivering the next bolus.
Bolus Too high
j
The bolus is above the
allowed amount. The bolus
was set to 50 units.
j
check the accuracy of all entries. If necessary,
contact your healthcare professional.
Bolus Too Low
j
The insulin pump cannot
deliver a 0.1 unit of multiwave
bolus.
j
The pump will adjust the bolus to 0.2 units.