Locate the two sidepiece fixing bars into the chair 
frame.  Push the sidepiece all the way in (Fig 10) 
and release the plunger by rotating it until the cap 
engages with the base (Fig 11).

 

 

 

 

 

 

15.

 

Ensure the plunger is engaged by moving the 
sidepiece in and out 

 the sidepiece should not 

move. 

16.

 

Tighten the hand screw to secure the sidepiece. 

17.

 

Repeat steps 12 to 16 to fit the other sidepiece. 

18.

 

Carry out a functional check (see section 14). 

 
 
 
 
 
 
 
 
 
 
 

8.

 

Chair adjustment 

The Configura Advance chair can be adjusted in the 
following areas: 

1.

 

Chair Width range

 

 The chair width has 4 widths 

that can be set, adjusted using the armrest locking 
plungers and hand screws.  The available width 
dimensions are: 

400 mm (16”)

 

450 mm 

(18”)

 

500 mm (20”)

 

550 mm (22”)

 

2.

 

Chair Depth range

 

 The chair depth can be 

adjusted in 9 increments from approximately 400 

mm (16”) to 5

50 mm (22

”). The depth is adjusted 

using a lever and slide mechanism. 

3.

 

Armrest Height range 

 The armrest height can be 

set at 

165mm (6.5”), 200mm (8”) 

or 

225mm (9”). 

The height is adjusted by changing the armrest 
bracket fitting position.

 

 

4.

 

Footplate Height range 

 

The

 

footplate height can 

be adjusted from 330mm 

(13”) to 533mm (21”).  

The height is adjusted using two plungers. 

Chair Width adjustment 

1.

 

Make sure a patient is not in the chair. 

2.

 

Make sure the brakes on all 3 locking castors are 
locked on (See section 13).   

3.

 

On one side of the chair, underneath the base of 
the sidepiece, locate the locking plunger (Fig 8, 
Item 1). 

4.

 

Pull the plunger out and rotate 90 degrees to lock 
the plunger open (Fig 9). 

5.

 

Loosen the hand screw (Fig 8, Item 2). 

6.

 

With two hands, pull or push the sidepiece until the 
desired measurement is achieved. The 
measurement can be read just inside the 
sidepiece at the front between the seat and the 
sidepiece (See Fig 12). 

7.

 

Pull the plunger out, rotate 90 degrees and let go 
to lock the plunger in position (Fig 11). 

8.

 

Move the sidepiece in and out to make sure that 
the plungers have locked the sidepiece in position. 

9.

 

Tighten the hand screw. 

10.

 

Adjust the opposite sidepiece in a similar manner 
to match the same measurement on both sides. 

Figure 9

 

Figure 10 

Figure 11

 

WARNING 

 
Adjustment MUST be carried out by suitably 
trained and qualified personnel. 
A clinical assessment and risk assessment MUST 
be carried out by suitably trained and 
qualified personnel before adjustments are 
made to the chair. 
Do not adjust the width or the depth of the 
chair when a patient or user is in the chair. 

 

 

 

 

Содержание Configura Advance Chair

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Страница 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Страница 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Страница 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Страница 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Страница 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Страница 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Страница 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Страница 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Страница 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Страница 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Страница 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Страница 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Страница 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Страница 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Страница 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Страница 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Страница 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Страница 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Страница 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Страница 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Страница 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Отзывы: