Problem 

Possible solution(s) 

The chair is not 
functioning (electric 
version). 

Ensure that the chair is plugged 
in.  If it is necessary to move the 
chair around, the chair will need 
to be plugged in for around 10 
hours to fully charge the battery. 

The chair does not 
function as expected 
(electric version). 

Reset the chair by pressing button 
2 to fully retract the leg rest, then 
button 3 to fully recline the chair.  

Alarm beeping 
(electric version). 

Connect the chair to the mains 
power supply to recharge the 
battery. 

Backrest will not 
recline (manual 

version). 

Ensure the backrest lockout slide 
is released 

 see section 10 of this 

manual. 

Leg rest will not raise 
or lower, alarm sounds 
when button pressed 

(electric version) 

Chair is in forward tilt position.  
Move chair to standard tilt 
position. 

Alarm beeping quickly 
when button pressed 
(electric version) 

Fault in actuator.  Check all 
connections. Contact Accora if 
problem persists. 

Battery box warning 
light solid yellow 

Battery charging 

 OK. 

Battery box warning 
light flashing yellow 

Fault while charging battery.  
Contact Accora. 

 
If the Configura Advance chair still does not function 
correctly after following through the troubleshooting 
procedure, please contact Accora for further advice. 

 
17.

 

Storage

 

 
For problem-free storage we recommend: 
 
1.

 

Disconnect the chair from the electric supply 

2.

 

Remove the accessories  

3.

 

Wrap the chair and accessories or cover them so 
that the coating and plastic parts are not 
damaged 

4.

 

Chair should be stored in a temperature between  
10°C to +25°C (50°F to +77°F) for the electric version 
and between 

-20°C to +50°C (-4°F to +122°F) for the 

manual version.

 

5.

 

Chair should be stored in a relative humidity (non-
condensing) between 30% and 75% 
 

18.

 

Daily inspection 

 

Daily visual inspection is strongly recommended and 
may be carried out by carer, user or another person.  
 
The following checks must be carried out: 

 
1.  Does the chair operate as per its intended purpose 

without unexpected noise or motion? 

2.  Are there any signs of abuse or excessive wear? 
3.  Is any fabric torn or damaged? Pay particular 

attention to the seat overlay. 

4.   Check that the overlay is secured to the frame with 

the press studs. 

5.  Are all fixtures and fittings tight and secure? 
6.  Does the chair frame appear stable and secure? 
7.  Are all accessories fitted in line with the accessory 

manufacturer or accessory supplier’s instructions?

 

8.  Are all the castor brakes in the locked position? 
9.  Are all electrical cables (including accessories, e.g. 

air pump) secured and routed to prevent 
damage? 

10. Is the area around, above and below chair clear of 

possible obstruction? 

11. Is the footplate securely attached? 
12. Are the plungers securing the sidepieces locked 

and secure? 

13. Is there any risk of entrapment or patient injury? 
14. Are any electrical cables pinched, crushed or 

damaged in any way? 

 
If any damage, performance issue or cause for 
concern is noted during this inspection the chair should 
be withdrawn from service and appropriate steps 
should be taken. 
 

19.

 

General maintenance

 

 

 

 
Repairs to the Configura Advance chair must be 
carried by suitably trained and qualified personnel. 
 
After any maintenance has taken place a functionality 

check must be carried out (see section 14). 
 

20.

 

Guarantee 

 

Model  

Configura Advance chair 

Warranty period 

2 years 

 

21.

 

Disposal 

 

In the event of the disposal of materials from the chair, 
end-of-life parts must be disposed of in accordance 
with current environmental regulations. 
Take care in disposing of batteries according to local 
regulations. 

WARNING 

 

Maintenance MUST be carried out by suitably 
trained and qualified personnel. 
All functions MUST be tested and approved 
after maintenance by suitably trained and 
qualified personnel.  
Only power supply supplied with chair may be 
used. 
Do not carry out maintenance with service 
user or patient on the chair.  
 

Содержание Configura Advance Chair

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Страница 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Страница 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Страница 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Страница 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Страница 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Страница 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Страница 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Страница 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Страница 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Страница 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Страница 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Страница 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Страница 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Страница 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Страница 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Страница 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Страница 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Страница 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Страница 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Страница 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Страница 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Отзывы: