background image

 

 

17 

14.

 

Functionality check 

 

These checks must be carried out after chair assembly, 
adjustment or maintenance. 
 
Using the manual levers or handset, test all chair 
functions. On the electric version, check all cables for 
risk of crushing. 
 

Manual version

 

1.

 

Check overlay poppers are engaged (Fig 7, Item1) 

2.

 

Tilt the chair backwards 

3.

 

Tilt the chair forward to upright position 

4.

 

Raise the leg rest 

5.

 

Lower the leg rest 

6.

 

Lower the backrest 

7.

 

Raise the backrest 

8.

 

Check the correct function of the castors. 

9.

 

Ensure footrest is fitted correctly. 

10.

 

Ensure sidepieces are fitted correctly. 

 

Electric version 

1.

 

Check overlay poppers are engaged (Fig 7, Item1) 

2.

 

Press button 3 to tilt the chair backwards. 

3.

 

Press button 4 to tilt the chair forwards to the 
upright position. 

4.

 

Press buttons 4 and 5 together to tilt the chair to 
the forward tilt position. 

5.

 

Press button 3 to return the chair to the upright 
position. 

6.

 

Press button 1 to raise the leg rest. 

7.

 

Press button 2 to lower the leg rest. 

8.

 

Check that all actions are completed smoothly 
and without any cables being crushed. 

9.

 

Check the correct function of the castors.  

10.

 

Ensure the handset cable is routed without any 
snags or kinked cable.  

11.

 

Ensure footrest is fitted securely. 

12.

 

Ensure sidepieces are fitted correctly. 
 
 
 
 
 
 

15.

 

Cleaning & disinfection 

 

 

Cleaning Information: 

To disinfect the chair, only use detergents designed for 
use in healthcare e.g. warm soapy water. Do not use 
abrasives, scourers or other materials that could 
damage the coating. Do not use corrosives, caustics or 
strong acids. 
Do not use detergents that could alter the structure or 
behaviour of the plastics (petrol etc.). 
 
Clean by wiping with a damp cloth.  Alternatively use 
Microfibre cleaning cloth without any soap. 
For extreme soiling use a 5% Sodium Hypochlorite 
solution. 
The chair is not designed for maintenance in 
automatic chair washers or for cleaning with 
pressurised water, spraying, showering or steam 
cleaning. 
Accora cannot be liable for any damage or risk of 
damage if inappropriate cleaning or disinfectant 
agents are used. 
 

Cleaning procedure: 

1.

 

Disconnect the chair from the power supply. 

2.

 

Move the chair to where cleaning will take place 
and lock the chair castors. 

3.

 

Remove the cushions and any accessories such as 
profiled headrest, external laterals. 

4.

 

Clean as described 

in the “Cleaning Information”.

 

 

16.

 

Troubleshooting 

 

 

WARNING 

 

Functionality check MUST be carried out by 
suitably trained and qualified personnel. 
Check for obstructions around, above and 
below the chair frame and position the chair 
so that it can operate through the full 
adjustment range without any chance of 
obstruction or entrapment.  
Always engage the castor brakes when the 
chair is stationary or left unattended. 

WARNING 

 

The chair must be disconnected from the 
power supply when being cleaned or 
disinfected. 
All functions MUST be tested and approved by 
a competent person after cleaning or 
disinfection. 
The chair MUST be cleaned and disinfected 
before re-using the chair for a different 
patient. 

WARNING 

 

Troubleshooting MUST be carried out by 
suitably trained and qualified personnel. 
Do not attempt to open any electrical part 
enclosures. 
Do not attempt to repair any electrical parts. 
All functions MUST be tested and approved by 
a competent person after troubleshooting. 
 
 

Содержание CHAIR-0-SC1-030

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Страница 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Страница 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Страница 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Страница 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Страница 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Страница 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Страница 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Страница 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Страница 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Страница 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Страница 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Страница 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Страница 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Страница 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Страница 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Страница 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Страница 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Страница 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Страница 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Страница 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Страница 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Отзывы: