17
14.
Functionality check
These checks must be carried out after chair assembly,
adjustment or maintenance.
Using the manual levers or handset, test all chair
functions. On the electric version, check all cables for
risk of crushing.
Manual version
1.
Check overlay poppers are engaged (Fig 7, Item1)
2.
Tilt the chair backwards
3.
Tilt the chair forward to upright position
4.
Raise the leg rest
5.
Lower the leg rest
6.
Lower the backrest
7.
Raise the backrest
8.
Check the correct function of the castors.
9.
Ensure footrest is fitted correctly.
10.
Ensure sidepieces are fitted correctly.
Electric version
1.
Check overlay poppers are engaged (Fig 7, Item1)
2.
Press button 3 to tilt the chair backwards.
3.
Press button 4 to tilt the chair forwards to the
upright position.
4.
Press buttons 4 and 5 together to tilt the chair to
the forward tilt position.
5.
Press button 3 to return the chair to the upright
position.
6.
Press button 1 to raise the leg rest.
7.
Press button 2 to lower the leg rest.
8.
Check that all actions are completed smoothly
and without any cables being crushed.
9.
Check the correct function of the castors.
10.
Ensure the handset cable is routed without any
snags or kinked cable.
11.
Ensure footrest is fitted securely.
12.
Ensure sidepieces are fitted correctly.
15.
Cleaning & disinfection
Cleaning Information:
To disinfect the chair, only use detergents designed for
use in healthcare e.g. warm soapy water. Do not use
abrasives, scourers or other materials that could
damage the coating. Do not use corrosives, caustics or
strong acids.
Do not use detergents that could alter the structure or
behaviour of the plastics (petrol etc.).
Clean by wiping with a damp cloth. Alternatively use
Microfibre cleaning cloth without any soap.
For extreme soiling use a 5% Sodium Hypochlorite
solution.
The chair is not designed for maintenance in
automatic chair washers or for cleaning with
pressurised water, spraying, showering or steam
cleaning.
Accora cannot be liable for any damage or risk of
damage if inappropriate cleaning or disinfectant
agents are used.
Cleaning procedure:
1.
Disconnect the chair from the power supply.
2.
Move the chair to where cleaning will take place
and lock the chair castors.
3.
Remove the cushions and any accessories such as
profiled headrest, external laterals.
4.
Clean as described
in the “Cleaning Information”.
16.
Troubleshooting
WARNING
Functionality check MUST be carried out by
suitably trained and qualified personnel.
Check for obstructions around, above and
below the chair frame and position the chair
so that it can operate through the full
adjustment range without any chance of
obstruction or entrapment.
Always engage the castor brakes when the
chair is stationary or left unattended.
WARNING
The chair must be disconnected from the
power supply when being cleaned or
disinfected.
All functions MUST be tested and approved by
a competent person after cleaning or
disinfection.
The chair MUST be cleaned and disinfected
before re-using the chair for a different
patient.
WARNING
Troubleshooting MUST be carried out by
suitably trained and qualified personnel.
Do not attempt to open any electrical part
enclosures.
Do not attempt to repair any electrical parts.
All functions MUST be tested and approved by
a competent person after troubleshooting.