background image

 

4

 

1.

 

Means of delivery 

 

The Configura Advance chair is supplied boxed.  An 
inspection must take place upon receipt to ensure the 
delivery is complete and undamaged. 
Any missing parts, faults or damage must be reported 
immediately to the carrier and Accora in writing. 

 

2.

 

Safety instructions 

 

1.

 

Before using the chair, you must read the 
instruction manual and use the chair in 
accordance with it. 

2.

 

The chair must not be used if faults have been 
detected on it that may injure the patient, staff or 
a third person. 

3.

 

The chair must only be operated by persons who 
are able to operate it in accordance with this 
manual. 

4.

 

Users must make the patient aware of the control 
functions that apply to the patient, subject to an 
assessment by a professional. 

5.

 

Before using the chair, the user should understand 
the chair and its functionality. 

6.

 

The safe working load, as specified in Section 4, 
must never be exceeded. 

7.

 

If a patient is left alone in the chair, the chair 
castors must be locked as an unlocked chair can 
cause injury to a patient who leaves the chair or 
changes position. 

8.

 

When operating the moving parts of the chair, 
care must be taken to ensure that the patient, 
other people and objects do not become 
trapped. 

9.

 

Before cleaning the chair, the electrical supply 
must be disconnected. 

10.

 

The chair may not be used where there is a 
danger of explosion or in the presence of 
uncontained flammable liquids. 

 

3.

 

Intended use 

 
1.

 

The Configura Advance chair is intended to be 
used as a portable chair with built in size, pressure 
and posture options in the acute, long-term care 
and homecare environments. The chairs are 
intended for patients who do not exceed the safe 
working load and height recommendations for the 
device. 

2.

 

The Configura Advance chair range is intended for 
use in primary healthcare environments where 
supervision can be provided if necessary, for 
example, long-term residential care, nursing homes 
or convalescent units. The Configura Advance 
chair may also be placed in private homes where 
there is unlikely to be 24-hour supervision by clinical 

staff, but where the patient or caregiver is 
considered competent in its use.  

3.

 

A typical patient / user ranges from those who are 
physically dependent on caregivers for their 
mobility, social and daily care needs, through to 
rehabilitating patients. 

4.

 

Patients using the chair may also have declining 
cognitive awareness, with symptoms ranging from 
mild forgetfulness and poor concentration through 
to end-stage dementia.  

5.

 

Patients may require postural support for the 
accommodation, encouragement and 
management of good posture. 

 

4.

 

Technical specification 

 
The table below lists the Configura Advance chair part 
numbers; some may not be available in your region. 
Part numbers may be suffixed (EU etc) to show regions. 

General specification 

Environmental information: 

Note: Always store out of direct sunlight.  If the chair is 
stored in conditions outside the normal operating 
range, it should be allowed time to stabilise, in normal 
operating conditions, before use. 

Description

 

Value 

Model name 

SC1, SC2, SC3, SC4 

Overall dimensions 

700mm W × 880mm D x 1300mm H 
27.6in W x 34.6in D x 51.2in H 

Chair castor 

4 x 100mm or 2x 125mm/2x 150mm 
with brake  

Safe Working Load 

160kg  

Maximum Patient 

Weight 

160kg / 352lbs / 25 st 

Audible noise 

<60 dBA 

Mass of Chair  

85.1kg / 188lbs 

Flammability 

Cover fabric: BS EN 7161(BS 

EN1021-1 and BS EN1021-2) 

Expected service life 

Typically 7 years 

Condition 

Temperature 

Range 

Relative 

Humidity 

Atmospheric 

pressure 

Operating 

+10°C to +25°C 
(+50°F to +77°F) 

30% to 75% 

(Non-

condensing) 

700 hPa to 

1060 hPa 

Transport/ 
storage -
Manual version 

-20°C to +50°C   
(-4°F to +122°F) 

Transport/ 
storage - 
Electric version 

+10°C to +25°C 
(+50°F to +77°F) 

Part number 

Description 

CHAIR-0-SC1-030 

Advance chair, 100mm castors, 
standard backrest, manual 

CHAIR-0-SC2-030 

Advance chair, 100mm castors, 

standard backrest, electric 

CHAIR-0-SC3-030 

Advance chair, 125/150mm 
castors, standard backrest, manual 

CHAIR-0-SC4-030 

Advance chair, 125/150mm 

castors, standard backrest, electric 

Содержание CHAIR-0-SC1-030

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Страница 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Страница 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Страница 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Страница 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Страница 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Страница 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Страница 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Страница 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Страница 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Страница 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Страница 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Страница 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Страница 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Страница 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Страница 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Страница 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Страница 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Страница 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Страница 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Страница 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Страница 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Отзывы: