AH-TX Series Handheld Pulse Oximeter Instruction Manual
50
Guidance and Declaration – Electromagnetic Immunity
The device is suitable for use in the electromagnetic environment specified
below. The customer or the user of the device should make certain that they
are in such an environment when operating it.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient / burst
(EFT)
IEC 61000-4-4
N/A
N/A
N/A
Surge IEC
61000-4-5
N/A
N/A
N/A
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
N/A
N/A
N/A