AH-TX Series Handheld Pulse Oximeter Instruction Manual
21
system for patient monitoring. Adjusting the alarm volume to a low
level may result in a hazard to the patient. Always keep the patient
under close surveillance.
General personnel should not arbitrarily modify the alarm
conditions, and only those who have professional knowledge or
relevant training can adjust the alarm setting.
4.5 When an Alarm Occurs
Warning:
When an alarm occurs, you should always check the patient’s
condition first.
Do not setting alarm limits to extreme values that can render the
alarm system useless.
Check the alarm message appeared on screen. It is needed to identify the
alarm and action appropriately, according to the cause of the alarm.
1.
Check the patient’s condition.
2.
Identify alarming parameter and alarm category.
3.
Identify the cause of the alarm.
4.
Silence the alarm, if necessary.
5.
When cause of alarm has been over, check that the alarm system is
working properly.
Alarm messages for individual parameters can be found in
Appendix D
Alarm message
.