AH-TX Series Handheld Pulse Oximeter Instruction Manual
20
When an alarm occurs, the alarm message will be displayed in the alarm area.
The system uses the following symbols to match the alarm level of
physiological alarm messages:
High level alarms: !!!
Medium level alarms: !!
Low level alarms: !
The system uses the following background colors to indicate different
messages and match the alarm level:
High level alarms(!!!): Red
Medium level alarms(!!): Yellow
Low level alarms(!): Yellow
4.4 Pausing the Alarm Tones
Press the alarm pause icon “
” to keep the alarm paused for 120 seconds
and the pause time will be displayed in red numeric.
When the audible alarm is paused, the alarm background color remains
lit, and the alarm message remains displayed.
Pressing the “
” icon again will restart the audible alarm.
The audible alarm automatically starts again once the alarm pause period
expires.
Warning:
When the alarm sound is switched off, the monitor will give no
audible alarm tones even if a new alarm occurs. Therefore, the user
should be very carefully about whether to switch off the alarm
sound or not.
Users should not rely exclusively on the audible alarm