4.3.2 Alarm Lamp
When an alarm occurs, the severity is indicated on the
alarm lamp in the following visual ways:
Alarm level
Visual prompt
High
Alarm lamp flashes in red at 2 Hz
Medium
Alarm lamp flashes in yellow at 0.5 Hz.
Low
Alarm lamp is yellow but does not flash.
Caution
:
When multiple alarms of different levels occur at
the same time, the oximeter will select the highest
warning level and give the highest visual and
audible alarm indications.
When multiple alarms occur at the same time, the
alarm messages will be displayed in the alarm area
in turn.