Handheld pulse oximeter user’s manual
5.6.4 Altering the Desaturation Limit
When the oximeter records an SpO
2
value that is lower
than the lower saturation limit, a high physiological alarm
will be triggered. The desaturation limit may be reset is as
follows:
1.
Select
【
Menu
】→【
System
】→【
Maintenance
】,
and a
pop up window will appear, requiring a password.
2.
Input the password and select
【
OK
】
to enter the
maintenance window. Select
【
Desat Lim.
】
, and then set
its value through the Up and Down button.
5.7 PR Alarm Setup
5.7.1 Switching PR Alarm On/Off
1.
Select
【
Menu
】→【
Alarm Setup
】
.
2.
Set the
【
Alarm
】
of PR to
【
Off
】
to shut off PR alarm.
When the PR alarm is, off, this symbol will be in the
visible PR parameter display area:
5.7.2 Setting the Alarm Level
1.
Select
【
Menu
】→【
Alarm Setup
】
.
2.
Set the
【
Alarm
】
of PR to
【
Med
】
or
【
High
】
.
7