Handheld pulse oximeter user’s manual
6.3 Reviewing Setup
After entering the reviewing window, press the left
button to enter the
【
Trend Setup
】
window, as shown below:
Fig 6-2 Trend Setup
In the window you can set
【
Interval
】
,
【
Select ID
】
,
【
Delete Selected
】
,
【
Delete All
】
and
【
Export Trend
】
:
z
Interval
:
The time interval between recordings
can be adjusted to take regular readings within the
range of 2 seconds to 30 minutes.
z
Select ID
:
This selects the desired patient ID No.
The user may change ID Nos. to browse the trend
data of corresponding patients.
z
Delete Selected
:
This deletes the stored trend data
of the selected ID No.