Quality
Assurance
Logs
Pentra
80 - User Manual - RAB108EA
3–35
◆
Every time that an analysis made on a QC is rejected, a notification is done.
5.4. Reagent logs
◆
For every reagent change, a notification is done.
5.5. Settings logs
◆
At every modification in the settings, a notification is done.
5.6. Maintenance logs
◆
For every maintenance, a notification shall be done by the user or technician in this screen.
After any maintenance operation, open the «Maintenance» logs screen. Press the «Insert»
key in order to enter an action, the duration and a comment about this action. Press «ok»
key to save the new entry.
5.7. Error logs
◆
A notification shall be done when an alarm is displayed
5.8. Host logs
◆
Each time a file from the Host is rejected, a notification is done. This rejection is generated
in the software rather than the raster when, for example:
– An order is pending a ReRun
– The rack on which the tube is found is in progress of an analysis
5.9. Blank cycle logs
For every Startup analysis performed, a notification is done.
◆
A Startup has passed if the cycle is launched and the WBC, RBC, HGB and PLT parameters
are within acceptable limits. the notification line is displayed in green.
◆
A Startup has failed if the cycle is launched and if the three consecutive analyses done
gave wrong results on the WBC or RBC or HGB or PLT parameters, the notification line is red.
◆
A Startup has failed if HGB blank value is not within acceptable limits. The notification
line is red.
◆
The default limit values are:
– WBC = 0.3 10
3
/ mm
3
– RBC = 0.03 10
6
/mm
3
– HGB = 0.3 g/dl
– PLT = 7 10
3
/mm
3
Содержание Pentra 80 RAB108EA
Страница 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Страница 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Страница 34: ...Pentra 80 1 22 Pentra 80 User Manual RAB108EA Fig 1 18 Order management Patient Sample Service Order LIS P80...
Страница 40: ...Pentra 80 1 28 Pentra 80 User Manual RAB108EA 6 Printer Use the printer supplied or approved by ABX DIAGNOSTICS...
Страница 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Страница 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Страница 176: ...Pentra 80 5 24 Pentra 80 User Manual RAB108EA 5 4 2 Printer properties Fig 5 22 Printer properties window...
Страница 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Страница 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Страница 256: ...Pentra 80 7 42 Pentra 80 User Manual RAB108EA 7 Hydraulic Diagram See Hydraulic diagram on next page...
Страница 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Страница 284: ......
Страница 286: ......
Страница 288: ......
Страница 290: ......
Страница 292: ......
Страница 294: ......
Страница 296: ......
Страница 297: ...Index Glossary Pentra 80 User Manual RAB108EA 9 1 Contents 1 Glossary 9 2 2 Index 9 4 Section9 Glossary...
Страница 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...