![ABLE Ultima OPIR-5 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/able/ultima-opir-5/ultima-opir-5_installation-and-maintenance-instructions-manual_3412579014.webp)
Ultima OPIR-5 Detector
xii
Alignment/Adjustment
Ensure that there is no significant amount of background gas when the unit is setting the zero
value, as this will alter the Ultima OPIR-5 Detector performance. If there is gas present, try
setting-up the Ultima OPIR-5 Detector on a breezy day, as this dissipates the gas.
1. It is recommended that two people are involved in aligning the Extended Range Unit.
One person should be at the Source and one person should be at the Receiver. An
alignment scope is helpful but not necessary. Otherwise follow the general Alignment and
Adjustment instructions.
2. At the Source, check the four screws that hold the unit to the pan-tilt assembly. Verify
that the back two screws have been tightened but that the front two have not. On the
pan-tilt assembly, verify that the front and back fine adjust screws are flush with the
assembly surface, to allow full movement of tilt. If necessary, use a 2.5 mm hex wrench
to adjust the screws. Use the housing sights to aim the unit at the center of the Receiver.
Align the notch and pin of the sights with the Receiver center.
3. At the Receiver, check the four screws that hold the unit to the pan-tilt assembly. Verify
that the back two screws have been tightened but that the front two have not. On the
pan-tilt assembly, verify that the front and back fine adjust screws are flush with the
assembly surface. Use the housing sights to aim the unit at the center of the Source.
Verify that the Receiver display is 0. Put the magnet on the upper right edge of the
display. The display will change from 0 to rst, ---, SE, and AJ (Adjustment). Remove the
magnet when AJ appears. The Receiver will display “A” with an AJ value from 0 to 99 or
“]-[“ if no Source signal is detected. Note that in 60 minutes, the alignment mode will time
Содержание Ultima OPIR-5
Страница 3: ...Ultima OPIR 5 Detector i This page intentionally left blank...
Страница 27: ...Ultima OPIR 5 Detector 12 Figure 7 Source Wiring Figure 8 Receiver Wiring...
Страница 52: ...Ultima OPIR 5 Detector 37 7 4 Engineering Documentation Figure 13 Outline Drawing...
Страница 55: ...MINE SAFETY APPLIANCES COMPANY CRANBERRY TOWNSHIP PENNSYLVANIA USA 16066 1 800 MSA INST www MSAsafety com...