5.39
Impella
®
System for Use During Cardiogenic Shock
3.
Confirm that the controller displays a pulsatile waveform and the inlet area of the
Impella
®
5.0 Catheter is 3.5 cm below the aortic valve.
Impella Flow
0.0
System Power
MUTE
ALARM
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella 5.0 S/N: 123456
30 - 06 - 2011 05:30
100%
CONTROL
FLOW
P - 0
L/min
0.0 Max
________________________
0.0 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 20.0 ml/hr
Motor
Current
0/0
( 0)
( mA)
10 sec.
Placement
Signal
( mmHg)
35/2
( 17 )
10 sec.
60
-20
1000
0
Figure 5.44 Pulsatile Waveform on Placement Screen
4.
Press the
FLOW CONTROL
soft button to open the P-level menu (see Figure 5.45).
5.
Turn the selector knob to increase the P-level from P-0 to P-2.
6.
Press the selector knob to select the new P-level.
Impella Flow
1.0
System Power
MUTE
ALARM
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella 5.0 S/N: 123456
30 - 06 - 2011 05:30
100%
CONTROL
FLOW
P - 2
L/min
1.2 Max
________________________
0.8 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 20.0 ml/hr
P - 2
P-7
P-8
P-5
P-6
P-9
P-2
P-3
P-0
P-1
P-4
Motor
Current
27 0/190
( 230)
( mA)
10 sec.
Placement
Signal
( mmHg)
35/2
( 17 )
10 sec.
60
-20
1000
0
Figure 5.45 Selecting P-level
7.
The catheter operation icon in the lower left corner of the screen begins rotating when
the Impella
®
5.0 Catheter beings to operate.
8.
Increase the P-level to P-9 to confirm correct and stable placement. Evaluate the
catheter position in the aortic arch and remove any excess slack. The catheter should
align against the lesser curvature of the aorta rather than the greater curvature. Verify
placement with fluoroscopy and with the placement screen (see Figure 5.46).
Positioning in Small
Hearts
If a patient has a smaller than
normal ventricular cavity, the
proper placement of the inlet
area of the catheter may be
3 cm (rather than 3.5 cm)
from the aortic valve.
Check Positioning at P-9
When the P-level is increased
to P-9, the Impella
®
5.0
Catheter has a tendency to
be drawn into the ventricle.
Check positioning at P-9 to
ensure proper placement
throughout the P-level setting
range.
5
U
S
IN
G
T
H
E A
U
TO
MA
TED
I
M
P
EL
LA
®
C
O
NT
R
O
LL
E
R
WI
TH
TH
E I
M
P
E
LL
A
®
C
A
TH
ETE
R
Содержание Impella 2.5
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Страница 16: ......
Страница 22: ......
Страница 38: ......
Страница 40: ......
Страница 108: ......
Страница 171: ......
Страница 173: ......
Страница 181: ......
Страница 183: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 210: ...INDEX TBD...