4.8
Instructions for Use & Clinical Reference Manual (US)
Table 4.3 Automated Impella
®
Controller Display Elements (continued)
Display Element
Description
Purge system area
(continued)
Y connector icon (for Impella
®
2.5 and Impella CP
®
)
•
Appears above the purge system marquee when the Impella
®
System is
configured using the Y connector in the set-up configuration
Purge flow
•
Current purge flow displayed in mL/hr below the purge system marquee
if the purge flow is known
•
Not displayed when the purge system is stabilizing, when there is no
purge cassette, or when the procedure has not yet started
Purge pressure
•
Current purge pressure (pressure of the purge fluid delivered through the
catheter to the motor) displayed in mmHg below the purge flow
Flow area
Information about Impella
®
Catheter flow is displayed in the lower left
corner of the display screen.
Max/Min
•
Max/Min displays the range for the flow rate
Current flow rate
•
Mean catheter flow displayed in liters per minute (L/min)—
the numbers appear in
white if the catheter position is correct; yellow if
the catheter position is incorrect or unknown
•
If the system is unable to calculate flow, a yellow triangular caution icon
is displayed with the message “Flow Calculation Disabled”
Catheter operation icon
•
The circular catheter operation icon rotates when the Impella
®
Catheter
is running
Central display area
On the home screen, the central display area displays a heart pictogram
and Impella
®
Catheter position indicator message.
Heart pictogram appears in the center of the home screen display.
•
Provides a visual representation of the current Impella
®
Catheter position
•
Overlaid with a translucent yellow “?” when the controller cannot
determine catheter position, position is wrong, and placement
monitoring is suspended or disabled
Impella
®
Catheter position indicator message displayed to the left of the
heart icon.
•
Displays “Impella Position OK” in green when catheter position is correct
•
Displays “Impella Position Unknown” in yellow when catheter position is
unknown
•
Displays “Impella Position in Ventricle” in yellow when catheter is in the
ventricle
•
Displays “Impella Position Wrong” in yellow when catheter position is
incorrect
•
Displays “Placement Monitoring Suspended” in yellow when there is a
fault in the sensor
•
Displays “Placement Monitoring Disabled” in yellow when you turn off
placement monitoring through the menu
Purge System
Stabilization
The purge system must
stabilize after case start,
a purge procedure, or
resolution of a purge alarm.
During this time, it may take
up to 3 minutes for purge
system information to display
on the screen.
Содержание Impella 2.5
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Страница 16: ......
Страница 22: ......
Страница 38: ......
Страница 40: ......
Страница 108: ......
Страница 171: ......
Страница 173: ......
Страница 181: ......
Страница 183: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 210: ...INDEX TBD...