7.7
Impella
®
System for Use During Cardiogenic Shock
a. Correct Impella
®
Catheter Position (TEE)
•
Catheter inlet area 3.5 cm below the aortic valve
•
Catheter outlet area well above the aortic valve
•
Catheter angled toward the left ventricular apex away
from the heart wall and not curled up or blocking the
mitral valve
b. Impella
®
Catheter Too Far into Left Ventricle (TEE)
•
Catheter inlet area more than 4 cm below the aortic valve
•
Catheter outlet area across or near the aortic valve
c. Impella
®
Catheter Inlet in Aorta (TEE)
•
Catheter inlet area in aorta or near the aortic valve
•
Catheter pigtail too close to the mitral valve
d. Impella
®
Catheter in Papillary Muscle (TEE)
•
Catheter pigtail in papillary muscle
•
Catheter inlet area more than 4 cm below the aortic valve or
lodged between papillary muscle and the myocardial wall
•
Catheter outlet area too close to the aortic valve
Figure 7.2 Transesophageal Echocardiographic (TEE) Illustrations of Impella
®
Catheter Position