89
The radio receiver in the device is using the same
frequency and bandwidth as the transmitter.
Cables and transducers:
Cables and transducers are not used during normal
use of the device nor while programming the device.
Quality of Service for Wireless Technology
Bluetooth
®
Smart wireless technology enables
communication between the generator and the patient
controller. The quality of the wireless communication
link varies depending on the use environment
(operating room, recovery room, and home
environment).
After the patient controller is paired with a generator,
the Bluetooth wireless technology symbol is visible on
the patient controller screen. When the Bluetooth
Smart wireless technology connection is not active,
the symbol appears dimmed.
The standard Bluetooth Smart wireless technology
data rate is 1 Mbit/s. The quality of service (QoS)
should allow wireless data to be transferred at a net
rate of 2.5 kB/sec. Each connection interval includes
a semi-duplex transmission with a required
acknowledge, a transmission latency in each direction
Содержание St. Jude Medical Patient Controller 3875