76
Table 10. Possible causes and solutions for potential
issues with opening the MRI Mode screen from
the Mode screen
Problem
Possible Cause Solution
Cannot access
the Mode
screen.
The IPG is not
connected to the
patient
controller.
Try connecting
to the IPG again.
Table 11. Troubleshooting messages for MRI mode
Message
Description
MRI is Not Advised
There may be a problem
with the implanted
lead(s). Contact your
clinician.
The generator is not in
MRI mode, and an MRI
scan cannot be performed.
MRI is Not Advised
The generator battery
voltage is too low.
The generator is not in
MRI mode, and an MRI
scan cannot be performed.