Compliance level: ± 8 kV contact; ± 15 kV air
Electromagnetic environment – guidance: Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Immunity test: Magnetic; IEC 61000-4-8
IEC 60601 test level: 30 A/m
Compliance level: 30 A/m
Electromagnetic environment – guidance: Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical domestic, commercial, or hospital environment.
Immunity test: Radiated RF; IEC 61000-4-3
IEC 60601 test level: 10 V/m; 80 MHz to 2.7 GHz
Compliance level: 10 V/m
Immunity test: Proximity fields from RF wireless communications equipment; IEC 61000-4-3
IEC 60601 test level: See table below
Compliance level: Compliance to the tested levels
Electromagnetic environment - guidance:
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the System, including cables
specified by Abbott Diabetes Care. Otherwise, degradation of the performance of the System could result.
The table below lists the immunity test levels at specific test frequencies for testing the effects of some
wireless communications equipment. The frequencies and services listed in the table are representative
examples in various locations where the System may be used.
DOC44383_rev-C 1:14pm 22-SEP-2021