day of use. 15 Sensors (14.9%) had “early Sensor shut-off” where the Sensor
algorithm detected that the Sensors did not function as intended and presented
the user with a Replace Sensor message. Table 7 displays the data for each day in
the wear duration for Study 3.
Table 7: Sensor Survival Rate Over Wear Duration (n=101)
Glucose Reading Availability
The System is designed to log a glucose reading every minute throughout the
wear period after the start-up time. Table 8 shows the glucose reading capture
rate for each day of the wear duration for Study 3.
Table 8: Glucose Reading Capture Rate Over Wear Duration (n=101)