Table 11. Guidance and Manufacturer's Declaration - Electromagnetic Emissions

Emissions Test

Compliance

Electromagnetic Environment

Guidance

Voltage fluctuations/ flicker emissions

IEC 61000-3-3

power supply network that supplies

buildings used for domestic purposes.

Class 5 Full Compliance*

*Only applicable to Power Adapter (Model 16720)

 

Table 12. Guidance and Manufacturer's Declaration - Electromagnetic Immunity

Immunity Test

IEC 60601

Test Level

Compliance Level

Electromagnetic

Environment Guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±8 kV contact

±15 kV air

±8 kV contact

±15 kV air

The relative humidity

should be at least 10%.

Electrical fast

transient/burst

IEC 61000-4-4*

±2 kV for power supply

lines

±1 kV for input/output lines

±2 kV for power supply

lines

±1 kV for input/output lines

Mains power quality should

be that of a typical

commercial, hospital, or

home environment.

Surge

IEC 61000-4-5*

±1 kV line to line

±2 kV line to ground

±1 kV line to line

±2 kV line to ground

Mains power quality should

be that of a typical

commercial, hospital, or

home environment.

Voltage dips, short

interruptions

IEC 61000-4-11*

<5% U

T

 (>95% dip in U

T

)

for 0.5 cycle

40% U

T

 (60% dip in U

T

)

for 5 cycles

70% U

T

 (30% dip in U

T

)

for 25 cycles

<5% U

T

 (>95% dip in U

T

)

for 5 s

<5% U

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

 (60% dip in U

T

)

for 5 cycles

70% U

T

 (30% dip in U

T

)

for 25 cycles

<5% U

T

 (>95% dip in U

T

)

for 5 s

Mains power quality should

be that of a typical

commercial, hospital, or

home environment. If the

user of the device requires

continued operation during

power mains interruptions,

it is recommended that the

device be powered from an

uninterruptible power

supply or a battery.

Power frequency

(50/60 Hz) magnetic field

IEC 61000-4-8

30 A/m

30 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical

commercial, hospital, or

home environment.

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to

80 MHz

3 Vrms

The device is intended for

use in the electromagnetic

environment in which

radiated RF disturbances

are controlled. The

customer or the user of the

device can help prevent

electromagnetic

interference by maintaining

a minimum distance

between portable and

mobile RF communications

equipment (transmitters)

and the device as

recommended below,

according to the maximum

output power of the

6 Vrms

ISM/Radio bands between

150 kHz to 80 MHz

80% AM at 1 kHz

6 Vrms

Radiated RF IEC 61000-4-3

10 V/m

80 MHz to 2.7 GHz

80% AM at 1 kHz

10 V/m

24

Содержание Eterna 16000

Страница 1: ...Charging System Eterna Spinal Cord Stimulation System Model 16000 User s Guide...

Страница 2: ...California to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a third party...

Страница 3: ...pparel 13 Traveling with Your Charging System 14 Troubleshooting 14 Troubleshooting Charging Issues 15 Service and Ordering Information 17 Appendix A Regulatory Statements 18 Disposal Guidelines for B...

Страница 4: ...ii...

Страница 5: ...e contraindications warnings precautions adverse effects and the personal identification card related to your neurostimulation system MRI Safety Information Do not bring your charger into the scanner...

Страница 6: ...it to dry completely before use Precautions The following precautions apply to this charging system Device components The charging system is for use with the Eterna implantable pulse generator only Y...

Страница 7: ...t site while charging your generator stop using the charger Use your patient controller to change the charging mode or contact your physician Heating or discomfort at surgical staples while charging C...

Страница 8: ...er your generator and charge the generator without having to manually hold the charger in place The lumbar apparel is worn like a belt around your waist Charging apparel pectoral Charging apparel is u...

Страница 9: ...the power icon blinks quickly Battery icon The battery icon illuminates in segments to indicate the charger battery level While the charger is being charged a segment of the battery icon blinks NOTE...

Страница 10: ...r to charge your generator To prepare the charger for pairing do the following 1 Ensure that your patient controller and charger are charged 1 2 Place the patient controller near the charger 2 3 Turn...

Страница 11: ...e or longevity may be reduced over time If your generator is depleted completely it will take a few minutes for the generator to turn on again and communicate with your patient controller once you sta...

Страница 12: ...ter 3 NOTE While the charger is charging it cannot be used to charge the generator If your charger is not charging you may be using an incorrect cable or power adapter Refer to the appendix for techni...

Страница 13: ...ly covering the charger 4 5 Wrap the charging apparel around your waist so the charger is over your generator 5 6 Adjust the velcro strap so the apparel is secure around your waist and the charger is...

Страница 14: ...the charger over your generator 2 3 Place the counterweight over your shoulder so it rests on your back or around your 3 neck so the counterweight rests on your opposite shoulder NOTE You may need to...

Страница 15: ...nsure that the charger is positioned over your generator 5 6 Fasten the buckle 6 NOTE If needed use the excess strap near the buckle to tighten or loosen the apparel Option 2 Loop and charger on same...

Страница 16: ...so that you can stop the charger if charging becomes uncomfortable To charge your generator follow these steps 1 Ensure the charger is not connected to an electrical outlet 1 NOTE If the charger is co...

Страница 17: ...charging is complete While the charger is charging the generator the charger cannot be turned off If you need to turn off the charger while it is charging the generator move it away from the generato...

Страница 18: ...your charging system follow these guidelines Use the travel case to store your charging system Be sure to pack the charging cable and power adapter NOTE For short trips if you plan to travel without...

Страница 19: ...he top of your generator Enable sound on the charger to use audible tones while aligning the charger over the generator A series of tones will indicate that charging has begun Avoid sitting on or agai...

Страница 20: ...ortable contact your physician The charger is not charging Ensure the charging cable is fully inserted and USB IF certified To determine if a cable is USB IF certified look for markings such as the fo...

Страница 21: ...ontroller differently Reconnect the patient controller with the charger Move to a different area or room away from interference in the environment or interference from other devices Avoid operating ot...

Страница 22: ...of FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules This device has been tested and found to comply...

Страница 23: ...format Table 5 Registration identification information Identifier Type Registration Identifier FCC Registration Number PX2 GMCHG1 Wireless Technology Information The following table summarizes technic...

Страница 24: ...th wireless technology symbol is visible on the clinician programmer or patient controller in the upper right corner of the screen When the Bluetooth wireless connection is not active the symbol appea...

Страница 25: ...communication from the charger to a legitimate patient controller or clinician programmer Troubleshooting for Wireless Coexistence or Interference Issues If you experience issues with interactions wi...

Страница 26: ...tion per the USB protocol Table 9 Implantable pulse generator compatibility for the charging system Description Model Eterna implantable pulse generator 32400 Appendix C Technical Specifications Use t...

Страница 27: ...evices contained in IEC 60601 1 2 2014 AMD1 2020 Ed 4 1 and CISPR11 2015 AMD1 2016 AMD2 2019 BS EN 55011 2016 A2 2021 However the device may cause radio interference or may disrupt the operation of ne...

Страница 28: ...5 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT...

Страница 29: ...ance and Manufacturer s Declaration Proximity Fields from RF Wireless Communication Equipment Proximity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance IEC 61000 4 3 38...

Страница 30: ...ic Fields Immunity Test Test Frequency Modulation Immunity Test Level A m IEC 61000 4 39 30 0 kHz Continuous wave 8 134 2 kHz Pulse modulation 2 1 kHz 65 13 56 MHz Pulse modulation 50 kHz 7 5 Appendix...

Страница 31: ...p away from rain Ingress protection rating for a device that is protected from the intrusion of solid foreign objects as small as 12 5 mm in diameter and is protected from vertically dripping water wh...

Страница 32: ...materials will help to conserve natural resources Authorized representative in the European Community European conformity affixed according to the relevant provisions of European Council Regulation 2...

Страница 33: ...Table 16 Symbols and definitions Symbol Definition This device is a recognized component by Underwriters Laboratories UL Inc This device complies with part 18 of the FCC rules 29...

Страница 34: ......

Страница 35: ......

Страница 36: ...abbott com 2023 05 ARTEN600283824 B 600283824...

Отзывы: