Table 11. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment
Guidance
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
power supply network that supplies
buildings used for domestic purposes.
Class 5 Full Compliance*
*Only applicable to Power Adapter (Model 16720)
Table 12. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
The relative humidity
should be at least 10%.
Electrical fast
transient/burst
IEC 61000-4-4*
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Surge
IEC 61000-4-5*
±1 kV line to line
±2 kV line to ground
±1 kV line to line
±2 kV line to ground
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Voltage dips, short
interruptions
IEC 61000-4-11*
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
Mains power quality should
be that of a typical
commercial, hospital, or
home environment. If the
user of the device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial, hospital, or
home environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
The device is intended for
use in the electromagnetic
environment in which
radiated RF disturbances
are controlled. The
customer or the user of the
device can help prevent
electromagnetic
interference by maintaining
a minimum distance
between portable and
mobile RF communications
equipment (transmitters)
and the device as
recommended below,
according to the maximum
output power of the
6 Vrms
ISM/Radio bands between
150 kHz to 80 MHz
80% AM at 1 kHz
6 Vrms
Radiated RF IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
10 V/m
24