Wireless Security Measures
The wireless signals are secured through device system design that includes the
following:
▪
The charger encrypts its wireless communication with the programming device using
▪
a key that is unique to that link.
▪
Only one patient controller or clinician programmer may communicate with the
▪
charger at a time.
▪
Standard Bluetooth® Low Energy (LE) pairing methods ensure valid and legitimate
▪
pairing among devices.
▪
Proprietary authentication for the patient controller or clinician programmer
▪
applications in addition to the pairing procedure specified in Bluetooth LE protocol,
which includes an element of proximity.
▪
A proprietary algorithm that detects and prevents an unauthorized application from
▪
communicating with the charger.
▪
Whitelisting methods that prevent unauthorized devices from using Bluetooth LE
▪
scanning to interfere with communication from the charger to a legitimate patient
controller or clinician programmer.
Troubleshooting for Wireless, Coexistence, or Interference
Issues
If you experience issues with interactions within the neurostimulation system, try the
following:
▪
Decrease the distance between the devices.
▪
▪
Move the devices so they share line of sight.
▪
▪
Move the devices away from other devices that may be causing interference.
▪
▪
Close the patient controller app; then turn the patient controller off and on.
▪
▪
Wait a few minutes and try connecting again.
▪
▪
Avoid operating other consumer electronic devices, such as a laptop, tablet, mobile
▪
phone, or cordless phone at the same time.
NOTE: Equipment such as the following can interfere with the neurostimulation
system: wireless home network devices, mobile and cordless telephones, and
consumer electronic devices.
Appendix B: Product Specifications
The charging system has the following product specifications and compatibility
information.
Table 8. Product specifications
Charger input
5 V, 0.5 A
Charger connector type
USB Type C
21
Содержание Eterna 16000
Страница 1: ...Charging System Eterna Spinal Cord Stimulation System Model 16000 User s Guide...
Страница 4: ...ii...
Страница 34: ......
Страница 35: ......
Страница 36: ...abbott com 2023 05 ARTEN600283824 B 600283824...