Table 12. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment Guidance
communications
equipment.
Recommended minimum
separation distance for
higher immunity test levels
where P is the maximum
power in watts (W), d is the
recommended separation
distance in meters (m), and
E is the immunity test level
in volts per meter (V/m).
*Only applicable to Power Adapter (Model 16720)
NOTE:
•
U
T
is the AC mains voltage prior to application of the test level.
•
•
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
•
reflection from structures, objects, and people.
Table 13. Guidance and Manufacturer's Declaration - Proximity Fields from RF
Wireless Communication Equipment
Proximity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment Guidance
IEC 61000-4-3
385 MHz: 27 V/m @ 18 Hz
pulse modulation
27 V/m
Recommended minimum
separation distance
d = 0.3 m
450 MHz: 28 V/m @ FM
modulation
28 V/m
710 MHz, 745 MHz,
780 MHz: 9 V/m @ 217 Hz
pulse modulation
9 V/m
810 MHz, 870 MHz,
930 MHz: 28 V/m @ 18 Hz
pulse modulation
28 V/m
1720 MHz, 1845 MHz,
1970 MHz: 28 V/m @
217 Hz pulse modulation
28 V/m
2450 MHz: 28 V/m @
217 Hz pulse modulation
28 V/m
5240 MHz, 5500 MHz,
5785 MHz: 9 V/m @ 217 Hz
pulse modulation
9 V/m
WARNING: Do not use portable RF communications equipment (including peripherals such as antenna cables
and external antennas) closer than 30 cm (12 inches) to any part of the device, including cables specified by
Abbott Medical. Otherwise, performance degradation may occur.
25