6
8.
Guide the rest of the lead into the Slitter tool thumbpad groove and secure it with
thumb.
CAUTION: Lead must be fully seated within the Slitter tool lead retention
channel and thumbpad groove.
CAUTION: Thumb position should not extend above thumbpad.
•
The lead fully seated in the lead retention
channel
•
Lead not seated fully in the retention channel
9.
Ensure the Slitter tool reference line is parallel with the catheter shaft (see figure
below).
Figure 2. Slitter Tool-Catheter Shaft Alignment
•
For a successful slit, the slitting reference
line on the Slitter tool must be parallel to
the catheter shaft.
•
Aligning the blade of the Slitter tool
perpendicular to the catheter shaft will
result in a difficult slit.