SensyMaster FMT230, FMT250
THERMAL MASS FLOWMETER | OI/FMT230/250-EN REV. C
89
Modbus register
address
Parameter name
Data type [register length] / value
range
Description
... / ...Alarm Simulation
4002 Alarm
Simulation
TUSIGN8 [1]
0: Aus
1: Mass flowrate exceeds limits.
2: Volume flowrate exceeds limits.
3: Simulation is on! Simulating
process/output value
4: Flowrate to zero
5: Maintenance interval is reached
6: All totalizer stopp.
7: Totalizer reset. Reset of one or more
Totalizers.
8: Display value is<1600h at Qmax.
9: Device not calibrated.
10: Sensor memory defective.
11: NV data defect. Data storage
irreparable.
12: No Frontend Board detected. Wrong
connection. Defect Frontend.
13: FEB communication error.
14: Incompatible Frontend Board.
15: NV chips defect on Motherboard.
16: Dig.Out 41/42 is saturated.
27: ADC Failure on Frontend Board.
28: Electronics failFrontend Board.
29: Sensor temperature out of range.
30: Frontend temp. out of range.
31: Sensor failure or disconnected.
32: Sensor heat emission limit.
33: Medium temperat exceeds limits.
34: Invalid Sensor configuration
35: Std.Volume flow exceeds limits.
36: Sensor soiling detected.
37: FEB voltages outside range.
38: Dig.Out 51/52 is saturated.
Manual simulation of alarms / error messages.
The simulated alarm is selected by setting the parameter to the
corresponding error number of the desired error.
See also chapter
Alarm status and alarm history status
on page 96.