SensyMaster FMT230, FMT250
THERMAL MASS FLOWMETER | OI/FMT230/250-EN REV. C
67
Parameter descriptions
Parameter range – Device Info
The parameterization of the device can be read out via the Modbus addresses listed here.
All Modbus addresses specified here are read only.
Modbus register
address
Parameter name
Data type [register length] /
value range
Description
… / Sensor
3421
Sensor Location Tag
TCHAR [20]
Sensor measuring point tagging
3401 Sensor
Tag
TCHAR [20]
Tag number of the sensor
2013
Sensor Element Type
TUSIGN8 [1]
0: Unknown
1: Standard
ceramics
2: Ceramics high temperature
design
3: Standard stainless steel
4: Hygienic stainless steel
5: Other
Sensor element type
2233 Sensor
Length
TFLOAT [2]
Installation length of the sensor
2012 Feature
Series
TUSIGN8 [1]
50: FMT230
60: FMT250
90: FMT410
100: FMT430
110: FMT450
Sensor model.
Specific functions such as the filling function are available depending
on the selection
3301 Sensor
ID
TUSIGN32 [2]
ID number of the sensor.
2501 Sensor
Serial
No.
TCHAR [20]
Serial number of the sensor.
3303
Sensor Run Hours
TUSIGN32 [2]
Operating hours of the sensor.
… / Sensor / Calibration
2016
First Cal. Date
TUSIGN8 [3]
Date of first calibration of sensor (calibration of new device).
2022
Last Cal. Date
TUSIGN8 [3]
Date of last calibration of sensor.
2521 Cal.
Cert.
No.
TCHAR [20]
Identification (number) of the relevant calibration certificate.
2541
First Cal. Location
TCHAR [20]
Place of first calibration of the sensor.
2561
Last Cal. Location
TCHAR [20]
Place of last calibration of sensor.
… / Sensor / ...Application Selector
6081 Application
TUSIGN8 [1]
1: Application
1
...
8: Application 8
Display of the selected application (type of measuring medium)