ScreenMaster RVG200
Paperless recorder
3 Installation
OI/RVG200-EN
13
Referring to Fig. 3.4, secure the recorder in a panel as follows:
a
Cut the correct sized hole in the panel – see Fig. 3.3, page 12.
b
Insert the recorder into the panel cut-out.
c
Fit upper and lower panel clamps.
d
Tighten the clamping screws evenly and securely to a torque of 20 Nm (177 lbf.in).
Caution.
Correct tightening of the clamping screws is critical in order to ensure proper
compression of the panel seal and achieve the IP66 / NEMA 4X hose-down rating.
Fig. 3.4 Installing the recorder
2
3
1
3
4
4
AC power supply earth
(ground) stud