111
FEM630 |
ELECTROMAGNETIC FLOWMETER | OI/FEM630-EN REV. A
9 . . .Diagnosis / error messages
. . .Error messages
Operation outside of specifications (Out Of Spec .)
Error no . / Range
Text on the LCD display
Cause
Remedy
S065.044 / operation
Inhouse temp. alarm.
Reduce ambient temperature.
Measuring medium or ambient temperature
is outside the spec.
Check process conditions, reduce
temperature.
S064.041 / operation
EPD alarm.
Secure pipe is completely filled.
Sensor not filled.
Check if pipe is empty.
Make sure that the sensor is completely filled.
S063.040 / operation
TFE alarm.
Secure pipe is completely filled.
Alarm of the complete filling electrode, but
incorrect, because the sensor is not
completely filled.
Check installation and process conditions.
S062.039 / operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
The measuring medium temperature is
outside the temperature limit.
Check process conditions and adjust alarm
threshold.
S061.038 / operation
Conductivity limits alarm.
Change limits or Check
application.
The conductivity of the measuring medium is
outside the limit values.
Check process conditions and adjust alarm
threshold.
S060.037 / operation
Gas bubble alarm.
Check conditions of application.
Gas bubbles in the measuring medium.
Check the process conditions.
S052.016 / operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
The selected process value of the current
output 31/32 is outside the measuring
range.
Adjust measuring range.
S051.017 / operation
Curr.Out V1 / V2, V3 / V4
saturated.
CO process value out of range.
Adapt Qmax.
The selected process value of the current
output V1 / V2 or V3 / V4 is outside the
measuring range.
Adjust measuring range.
S049.019 / electronics
Option Card 1 com error.
Defective Card. Check Card 1.
Call Service.
Broken hardware of the motherboard or
option card.
EMC interference
Check / replace option card in slot 1.
Please contact the service department.
S048.020 / electronics
Option Card 2 com error.
Defective Card. Check Card 2.
Call Service.
Broken hardware of the motherboard or
option card.
EMC interference
Check / replace option card in slot 2.
Please contact the service department.
S047.015 / operation
Pulse output is cut off.
Wrong config. Check pulse out
configuration.
The calculated output pulse or the calculated
output frequency is above the configured
cutoff frequency.
Check configuration for the output pulse.
S046.000 / operation
Mass flowrate exceeds limits.
Check flowrate and alarm limits.
The mass flow is below or above the
configured limit values ‘Qm Massflow Min’
and ‘Qm Massflow Max’.
Check the parameterization in menu ‘Process
Alarm / ...Alarm Limits’ and adjust as needed.
Check volume flow rate.
S044.001 / operation
Volume flowrate exceeds limits.
Check flowrate and alarm limits.
The volume flow rate is below or above the
configured limit values ‘Qv Volumeflow Min’
and ‘Qv Volumeflow Max’.
Check the parameterization in menu ‘Process
Alarm / ...Alarm Limits’ and adjust as needed.
Check volume flow rate.