9 . . .Diagnosis / error messages
. . .Overview
Error no . / Range
Error text
Current output
Digital output
Pulse output
LCD display
Error maskable?
S065.044 / operation
Inhouse temp. alarm.
Reduce ambient
temperature.
Current value - no
change.
No Answer
No Answer
Current value -
no change.
S064.041 / operation
EPD alarm.
Secure pipe is
completely filled.
Alarm - configured as in
menu ‘Curr. at EPD
Alarm’.
Alarm, if DO as ‘Logic /
Alarm Signal / EPD
Alarm’ is configured
0 Hz
0 %
Menu ‘...Group
Masking’.
S063.040 / operation
TFE alarm.
Secure pipe is
completely filled.
Alarm - configured as in
menu ‘Curr. at TFE
Alarm’.
Alarm, if DO as ‘Logic /
Alarm Signal / TFE
Alarm’ is configured
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S062.039 / operation
Sensor temp. limits
alarm.
Change limits or
change fluid
temperature.
Current value - no
change.
No Answer
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S061.038 / operation
Conductivity limits
alarm.
Change limits or
Check application.
Current value - no
change.
Alarm, if DO as ‘Logic /
Alarm Signal /
Conductivity’ is
configured
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S060.037 / operation
Gas bubble alarm.
Check conditions of
application.
Current value - no
change.
Alarm, if DO as ‘Logic /
Alarm Signal / Gas
bubble Alarm’ is
configured
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S052.016 / operation
Curr.Out 31 / 32 is
saturated.
CO process value out
of range. Adapt
Qmax.
Alarm - configured as in
menu ‘„Curr.Out >
20.5mA’.
Current value - no
change.
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S051.017 / operation
Curr.Out V1 / V2, V3
/ V4 saturated.
CO process value out
of range. Adapt
Qmax.
S049.019 /
electronics
Option Card 1 com
error.
Defective Card.
Check Card 1. Call
Service.
Does not react anymore Current value - no
change.
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.
S048.020 /
electronics
Option Card 2 com
error.
Defective Card.
Check Card 2. Call
Service.
Does not react anymore Current value - no
change.
Current value - no
change.
Current value -
no change.
Menu ‘...Group
Masking’.