Navigator 500
Sodium analyzer
8 Maintenance
62
OI/ASO550–EN Rev. G
4. If the
Calibrate
page is not displayed, use the
/
keys to scroll to the
Calibrate
page:
5. Press the
key (below the
Select
prompt).
–
If one or more single-stream wet-sections are
connected, the
Calibrate
/
Sodium 1 (2, 3, 4)
page is
displayed:
–
If a multi-stream wet-section is connected, the
Calibrate
/
Sodium Three Stream
page is displayed:
6. Press the
key (below the
Select
prompt).
The following page is displayed with the
Sensor
Calibration
menu selected:
7. Press the
key (below the
Select
prompt).
The
Sensor Calibration
page is displayed:
Press the
key to highlight the
Regeneration Cycle
menu.
8. Press the
key (below the
Select
prompt).
The
Regeneration Cycle
page is displayed:
Press the
key (below the
Yes
prompt) to start the
regeneration procedure.
Menu
Calibrate
Exit
Select
Calibrate
Sodium 1
Back
Select
Sodium Three Stream
Calibrate
Back
Select
Sensor 1
Sensor Calibration
Scheduled Calibration
Calibration Setup
Restore Defaults
Back
Select
Sensor Calibration
Low Calibration
High Calibration
Two-Point Cal
Regeneration Cycle
Process Calibration
Back
Select
Regeneration Cycle
No
Yes
Start Procedure